Increasing spatial resolution of 3T MRI scanning improves reproducibility of carotid arterial wall dimension measurements.

Magma 2014 June
OBJECTIVE: To improve carotid 3T magnetic resonance imaging (MRI) dimension measurements in patients with overt atherosclerotic carotid artery disease.

MATERIALS AND METHODS: In 31 patients with advanced atherosclerotic carotid artery disease, two high resolution (0.25 × 0.25 mm(2); HR) and two routinely used low resolution (0.50 × 0.50 mm(2); LR) carotid 3T MRI scans were performed within 1 month. After manual delineation of carotid wall contours in a dedicated image analyses program in eight slices covering the atherosclerotic plaque, image reproducibility, as well as the within-reader and between-reader variability were determined.

RESULTS: We found significantly higher intraclass correlation coefficients for total wall volume, mean wall area and mean wall thickness for the HR measurements (all p < 0.05). We found a significant lower signal-to-noise and contrast-to-noise ratio for the HR compared to the LR measurements. The carotid arterial wall dimension measurements of all parameters were significantly lower for the HR compared to the LR measurements. No significant differences were observed between the within-reader and between-reader reproducibility for HR versus LR measurements.

CONCLUSION: Increasing the in-plane resolution improves the reproducibility of 3T MRI carotid arterial wall dimension measurements. The use of HR imaging will contribute to a reduced sample size needed in intervention trials using MRI scanning of the carotid artery as surrogate marker for atherosclerosis progression.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app