Ability of the third-generation FloTrac/Vigileo software to track changes in cardiac output in cardiac surgery patients: a polar plot approach.

OBJECTIVE: To evaluate the ability of the third-generation (3.01) of FloTrac/Vigileo monitor (Edwards Lifesciences, Irvine, CA) to follow variations in cardiac output (∆CO) using the new polar plot approach.

DESIGN: Prospective interventional study.

SETTING: Single hospital university study.

PARTICIPANTS: Twenty-five patients referred for cardiac surgery.

INTERVENTIONS: CO was measured simultaneously by 3 to 5 bolus thermodilution (COtd measurements), using a pulmonary artery catheter and by arterial pulse contour analysis, using the FloTrac/Vigileo (COvi). Data were collected at eight time points: before incision, after sternotomy, before and after protamine sulfate infusion, at the start of sternal closure, at the end of surgery, on arrival to intensive care unit, and after a standardized volume expansion with 500 mL of hetastarch 6%.

MEASUREMENTS AND MAIN RESULTS: One-hundred thirty-five pairs of CO data were collected; the mean bias of all CO measurements corrected for repeated measures was 0.2 L/min with limits of agreements of -3.3 L/min and +2.9 L/min. The percentage error was 66.5%. The polar plot analysis included 71 significant ∆CO and showed a mean polar angle of -3.4 degrees with 95% polar percentage error equivalent limits of -61 to 55; 69% of analysed data points fell within the 30-degree limits and provided a correct polar concordance rate.

CONCLUSIONS: Third-generation FloTrac/Vigileo software still lacks the accuracy to reliably detect changes in cardiac output (∆CO) in cardiac surgery. Improvements to FloTrac/Vigileo CO algorithm and software still are needed in this particular setting.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app