Comparative Study
Journal Article
Multicenter Study
Randomized Controlled Trial
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Improved safety and reduction in stent thrombosis associated with biodegradable polymer-based biolimus-eluting stents versus durable polymer-based sirolimus-eluting stents in patients with coronary artery disease: final 5-year report of the LEADERS (Limus Eluted From A Durable Versus ERodable Stent Coating) randomized, noninferiority trial.

OBJECTIVES: This study sought to report the final 5 years follow-up of the landmark LEADERS (Limus Eluted From A Durable Versus ERodable Stent Coating) trial.

BACKGROUND: The LEADERS trial is the first randomized study to evaluate biodegradable polymer-based drug-eluting stents (DES) against durable polymer DES.

METHODS: The LEADERS trial was a 10-center, assessor-blind, noninferiority, "all-comers" trial (N = 1,707). All patients were centrally randomized to treatment with either biodegradable polymer biolimus-eluting stents (BES) (n = 857) or durable polymer sirolimus-eluting stents (SES) (n = 850). The primary endpoint was a composite of cardiac death, myocardial infarction (MI), or clinically indicated target vessel revascularization within 9 months. Secondary endpoints included extending the primary endpoint to 5 years and stent thrombosis (ST) (Academic Research Consortium definition). Analysis was by intention to treat.

RESULTS: At 5 years, the BES was noninferior to SES for the primary endpoint (186 [22.3%] vs. 216 [26.1%], rate ratio [RR]: 0.83 [95% confidence interval (CI): 0.68 to 1.02], p for noninferiority <0.0001, p for superiority = 0.069). The BES was associated with a significant reduction in the more comprehensive patient-orientated composite endpoint of all-cause death, any MI, and all-cause revascularization (297 [35.1%] vs. 339 [40.4%], RR: 0.84 [95% CI: 0.71 to 0.98], p for superiority = 0.023). A significant reduction in very late definite ST from 1 to 5 years was evident with the BES (n = 5 [0.7%] vs. n = 19 [2.5%], RR: 0.26 [95% CI: 0.10 to 0.68], p = 0.003), corresponding to a significant reduction in ST-associated clinical events (primary endpoint) over the same time period (n = 3 of 749 vs. n = 14 of 738, RR: 0.20 [95% CI: 0.06 to 0.71], p = 0.005).

CONCLUSIONS: The safety benefit of the biodegradable polymer BES, compared with the durable polymer SES, was related to a significant reduction in very late ST (>1 year) and associated composite clinical outcomes. (Limus Eluted From A Durable Versus ERodable Stent Coating [LEADERS] trial; NCT00389220).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app