We have located links that may give you full text access.
Raster-scan optoacoustic mesoscopy in the 25-125 MHz range.
Optics Letters 2013 July 16
We developed a raster-scan acoustic resolution broadband optoacoustic mesoscopy system and investigated the imaging performance using ultrasonic frequencies up to 125 MHz. The developed system achieves 7 μm axial resolution and transverse resolution of 30 μm reaching depths of at least 5 mm. This unprecedented performance is achieved by operating at out-of-focus ultrasonic detection and tomographic reconstruction. We demonstrate the limits reached due to the width of the laser pulse employed and showcase the technique on drosophila fly and drosophila pupae ex vivo.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app