Evaluation of a fully covered self-expanding metal stent with flared ends in malignant biliary obstruction: a multicenter study.

BACKGROUND AND AIMS: Limited data are available regarding fully covered metal stents in the management of malignant distal biliary strictures. The aim of this study was to evaluate the safety of a fully covered self-expanding metal stent (FCSEMS) with flared ends, in treating malignant biliary strictures. We report our long-term retrospective analysis from 6 centers.

METHODS: A total of 260 patients (142 males, median age 68 y) underwent endoscopic retrograde cholangiopancreatography with placement of FCSEMS (WallFlex; Boston Scientific, Natick, MA) for the palliation of obstructive jaundice in the setting of pancreatic adenocarcinoma (169), metastatic disease (36), cholangiocarcinoma (23), ampullary cancer (19), or other (13). Patients were evaluated clinically in follow-up and biochemical tests of liver function were obtained. Data were recorded for the following variables: patient survival, duration of stent patency, the need for subsequent biliary intervention, and complications.

RESULTS: A total of 266 FCSEMS were placed in 260 patients. There was a median survival of 100 days (range, 7 to 531 d). There was a mean follow-up of 134±118 days (range, 4 to 519 d). Biliary decompression was successful in 252 patients (97%). At the end of the study period, 121 were alive with a patent stent, 65 patients died with a patent stent, 40 patients were successfully bridged to surgery, 8 patients had their patent stent removed and had no need for further stenting, and 18 patients were lost to follow-up. Five patients who had a successfully placed FCSEMS were considered a failure due to the following reason: migration (2), cholangitis (1), stent occlusion (1), and removal for management of proximal biliary obstruction (1). Two additional patients experienced migration that did not require FCSEMS removal or replacement. Stent replacement was required in 10 patients, of whom 6 had a second FCSEMS placed. The remaining 4 were in the failure group and underwent replacement with either uncovered stents or plastic stents. Other complications, managed conservatively, included pain (2), postendoscopic retrograde cholangiopancreatography pancreatitis (4), fever (1), retroperitoneal perforation (1), and postsphincterotomy bleeding (1). The mean patency duration was 328 days (SE 19.04). The patency percentage was 83% at 3 months, 63% at 6 months, and 48.5% at 12 months.

CONCLUSIONS: In the management of malignant distal biliary strictures, the fully covered WallFlex stent has acceptable patency and complication rates. Further long-term prospective data are required to confirm this observation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app