Raw-data-based iterative reconstruction versus filtered back projection: image quality of low-dose chest computed tomography examinations in 87 patients.

Clinical Imaging 2013 November
OBJECTIVE: The objective was to compare standard-dose chest computed tomography (CT) reconstructed with filtered back projection (FBP) versus low-dose images with FBP and raw-data-based iterative reconstruction.

METHODS: Eighty-seven consecutive patients (46 male; mean age, 54.54 ± 16.12; mean body mass index, 24.58 ± 4.07) referred for initial chest CT with full-dose examinations [mean dose-length product (DLP), 183.37 ± 44.13 mGy · cm] and follow-up chest CT with half-dose examinations (mean DLP, 91.08 ± 23.81 mGy · cm) were included. The full-dose protocol was reconstructed with FBP; the half-dose protocol was reconstructed with FBP and sinogram-affirmed iterative reconstruction (SAFIRE). Noise and signal-to-noise ratio were compared using a paired Student's t test; subjective image quality and lesion conspicuity were compared using Wilcoxon signed ranks test.

RESULTS: Actual radiation dose of follow-up CT was about 50% (49.26% ± 2.62%) of standard-dose protocol. Compared to full-dose images with FBP, there was no significant difference in half-dose images with SAFIRE in the objective noise (ascending aorta: P=.38, descending aorta: P=.70, trachea on mediastinal images: P=.37) and SNR (ascending aorta: P=.14, descending aorta: P=.72, trachea on mediastinal images: P=.06) on mediastinal images. Noise was significantly lower (P<.001) and SNR was significantly higher (P<.001) in half-dose images with SAFIRE on lung images. Noise was significantly higher (P<.001) and SNR was significantly lower (P<.001) in half-dose images with FBP. Subjective image quality was similar on both mediastinal images (P=.317) and lung images (P=.614) of half-dose SAFIRE images versus full-dose FBP images. Lesion conspicuity was also similar. Subjective image quality was significantly lower on both mediastinal images (P<.001) and lung images (P<.001) of half-dose FBP images versus full-dose FBP images. The conspicuity of some lesions was significantly lower (ground-glass opacity, P<.0001; ill-defined micronodule, P<.0001; lung cyst, P<.0001; emphysematous lesion, P=.003) on half-dose FBP versus full-dose FBP images.

CONCLUSION: Compared to full-dose CT images reconstructed with the conventional FBP algorithm, SAFIRE with three iterations could provide similar or better image quality at 50% less dose.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app