Whether noninvasive optimization of AV and VV delays improves the response to cardiac resynchronization therapy

Bożena Urbanek, Michał Chudzik, Artur Klimczak, Marcin Rosiak, Joanna Lewek, Jerzy Krzysztof Wranicz
Cardiology Journal 2013, 20 (4): 411-7
23913460

BACKGROUND: Device optimization is not routinely performed in patients who underwent cardiac resynchronization therapy (CRT) device implantation. Noninvasive optimization of CRT devices by measurement of cardiac output (CO) can be used as a simple method to assess ventricular systolic performance. The aim of this study was to assess whether optimization of atrioventricular (AV) and interventricular (VV) delay can improve hemodynamic response to CRT and whether this optimization should be performed for each patient individually.

METHODS: Twenty patients with advanced heart failure New York Heart Association (NYHA) class III/IV, left ventricular ejection fraction ≤ 35% and left bundle branch block (QRS ≥ 120 ms) in sinus rhythm were evaluated from 24 h to 48 h after implantation of a CRT device by means of impedance cardiography (ICG). CO was first measured at each patient's intrinsic rhythm. Patients then underwent adjustments of AV and VV delay from 80 ms to 140 ms and from -60 ms to +60 ms, respectively in 20 ms increment steps and CO at each setting was measured by ICG. Both AV and VV delays were programmed according to the greatest improvement in CO compared to intrinsic rhythm.

RESULTS: There was a statistically signifi cant increase in CO measured at the intrinsic rhythm compared to different AV delay by mean of 21% (3.8 ± 1.0 vs. 4.6 ± 0.1 L/min, p < 0.05). Optimal AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing caused additional increased CO from intrinsic rhythm by mean of 32.6% (3.8 ± 1.0 vs. 5.04 ± ± 1.0 L/min, p < 0.05). Optimal AV/VV setting delays also resulted in improved hemodynamic responses compared to VV factory setting delay.

CONCLUSIONS: Both AV and VV delay optimization should be performed in clinical practice. Optimal AV delay improved outcome. However, combination of optimized AV/VV delays provided the best hemodynamic response. Optimized AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing increased hemodynamic output compared to intrinsic rhythm and VV factory setting delay.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23913460
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"