Inter- and intraobserver reliability of two-dimensional CT scan for total knee arthroplasty component malrotation

Beau Konigsberg, Ryan Hess, Curtis Hartman, Lynette Smith, Kevin L Garvin
Clinical Orthopaedics and related Research 2014, 472 (1): 212-7
23813180

BACKGROUND: Rotational malalignment of total knee arthroplasty (TKA) has been correlated with patellofemoral maltracking, knee instability, and stiffness. CT is the most accurate method to assess rotational alignment of prosthetic components after TKA, but inter- and intraobserver reliability of CT scans for this use has not been well documented.

QUESTIONS/PURPOSES: The objective of this study was to determine the inter- and intraobserver reliability and the repeatability of the measurement of TKA component rotation using two-dimensional CT scans.

METHODS: Fifty-two CT scans of TKAs being evaluated for revision surgery were measured by three different physicians. An orthopaedic resident and attending measured the same scans twice (more than 2 weeks apart) and a musculoskeletal radiologist measured them once. To assess interobserver reliability, intraclass correlation coefficients (ICCs) with two-way mixed-effects analysis of variance models as well as 95% confidence intervals for each were done. The repeatability coefficient was calculated as well, which is defined as the difference in measurements that include 95% of the values. This indicates the magnitude of variability among measurements in the same scale, which in this study is degrees.

RESULTS: The interobserver ICC measurement for the femoral component was 0.386 (poor), and it was 0.670 (good) for the tibial component. The interobserver ICC for the combined rotation measurements was 0.617 (good). The intraobserver ICC for the femoral component was 0.606 (good), and it was 0.809 (very good) for the tibial component. The intraobserver ICC for combined rotation was 0.751 (good). The intraobserver repeatability coefficient for the femoral component was 0.49°, 10.64° for the tibial component, and 12.29° for combined rotation.

CONCLUSIONS: In this study, the inter- and intraobserver reliability, and the repeatability, of TKA component rotation were variable. This raises concern about whether CT scan is diagnostic in the assessment of component malrotation after TKA.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23813180
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"