OBJECTIVES: This study sought to determine if the risk of mortality associated with inappropriate implantable cardioverter-defibrillator (ICD) shocks is due to the underlying arrhythmia or the shock itself.

BACKGROUND: Shocks delivered from ICDs are associated with an increased risk of mortality. It is unknown if all patients who experience inappropriate ICD shocks have an increased risk of death.

METHODS: We evaluated survival outcomes in patients with an ICD and a cardiac resynchronization therapy defibrillator enrolled in the LATITUDE remote monitoring system (Boston Scientific Corp., Natick, Massachusetts) through January 1, 2010. First shock episode rhythms from 3,809 patients who acutely survived the initial shock were adjudicated by 7 electrophysiologists. Patients with a shock were matched to patients without a shock (n = 3,630) by age at implant, implant year, sex, and device type.

RESULTS: The mean age of the study group was 64 ± 13 years, and 78% were male. Compared with no shock, there was an increased rate of mortality in those who received their first shock for monomorphic ventricular tachycardia (hazard ratio [HR]: 1.65, p < 0.0001), ventricular fibrillation/polymorphic ventricular tachycardia (HR: 2.10, p < 0.0001), and atrial fibrillation/flutter (HR: 1.61, p = 0.003). In contrast, mortality after first shocks due to sinus tachycardia and supraventricular tachycardia (HR: 0.97, p = 0.86) and noise/artifact/oversensing (HR: 0.91, p = 0.76) was comparable to that in patients without a shock.

CONCLUSIONS: Compared with no shock, those who received their first shock for ventricular rhythms and atrial fibrillation had an increased risk of death. There was no significant difference in survival after inappropriate shocks for sinus tachycardia or noise/artifact/oversensing. In this study, the adverse prognosis after first shock appears to be more related to the underlying arrhythmia than to an adverse effect from the shock itself.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23810882
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"