Comparison of 2 techniques used to obtain sterile urine specimens from urethral catheters.
Clinical Nurse Specialist CNS 2013 July
OBJECTIVE: The objective of this study was to compare the rates of urine contamination and the frequencies of minor adverse events between the vacuum tube and injector techniques methods for the collection of sterile urine specimens.
DESIGN: This study was a cross-sectional study.
SAMPLE: The sample population was composed of patients who underwent a urodynamics procedure.
SETTING: This study was performed at the urodynamics unit of a rehabilitation and care center in Turkey.
METHODS: Sterile urine specimens (n = 576) were collected from 144 patients using the vacuum tube technique and the injector technique; the rates of urine contamination and minor adverse events were evaluated. χ2 Statistics and Fisher exact tests were used to assess the contamination rates and frequencies of minor adverse events during the collection of sterile urine specimens.
RESULTS: The contamination rate of urine specimens collected using an injector was 0.9% (P > .05). The rate of minor adverse events during specimen collection was lower when using the vacuum technique than when using an injector and sterile urine container (2 vs 36 events, P < .05).
CONCLUSION: The vacuum tube technique of urine specimen collection was not superior to the injector technique in regard to contamination. The use of the vacuum tube technique for the collection of sterile urine specimens can prevent minor adverse events related to spillage/leakage of urine.
IMPLICATIONS: More research from multiple settings and work environments is needed to strengthen the evidence base of vacuum tube technique of sterile urine specimen collection.
DESIGN: This study was a cross-sectional study.
SAMPLE: The sample population was composed of patients who underwent a urodynamics procedure.
SETTING: This study was performed at the urodynamics unit of a rehabilitation and care center in Turkey.
METHODS: Sterile urine specimens (n = 576) were collected from 144 patients using the vacuum tube technique and the injector technique; the rates of urine contamination and minor adverse events were evaluated. χ2 Statistics and Fisher exact tests were used to assess the contamination rates and frequencies of minor adverse events during the collection of sterile urine specimens.
RESULTS: The contamination rate of urine specimens collected using an injector was 0.9% (P > .05). The rate of minor adverse events during specimen collection was lower when using the vacuum technique than when using an injector and sterile urine container (2 vs 36 events, P < .05).
CONCLUSION: The vacuum tube technique of urine specimen collection was not superior to the injector technique in regard to contamination. The use of the vacuum tube technique for the collection of sterile urine specimens can prevent minor adverse events related to spillage/leakage of urine.
IMPLICATIONS: More research from multiple settings and work environments is needed to strengthen the evidence base of vacuum tube technique of sterile urine specimen collection.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app