OBJECTIVES: To describe the initial world experience and mid-term follow-up of perimembranous ventricular septal defect (pmVSD) closure with a newly designed occluder.

BACKGROUND: Transcatheter closure of pmVSDs has been associated with a substantial risk of complete heart block, prompting many centers to abandon this intervention.

METHODS: A prospective multicenter cohort study was conducted on patients with pmVSD undergoing catheter closure using the Amplatzer(®) Membranous VSD Occluder 2 in the initial 4 pilot centers.

RESULTS: Nineteen patients, median age 6 years (range 1.4-62 years), were enrolled and followed for 14 ± 3 months (range 8-20 months). The median weight was 26 kg (range 9.3-96 kg) and the mean Qp/Qs ratio was 1.8 ± 0.7. The defect on left ventricular side measured 9.9 ± 3.5 mm and the orifice on right ventricular side was 8.1 ± 2.8 mm by echocardiography. Mean device size was 9.4 ± 2.4 mm (range 5-14 mm). An eccentric device was employed in 9 patients (47%) and a concentric device in 10 (53%). Overall, 18 patients (95%) had successful device implants. Procedural time was 122 ± 39 min. There were no procedural complications. Mild residual shunting was initially observed in 14 (78%) patients. At last follow-up, mild residual shunting persisted in only 3 (17%) patients. There was no significant increase in aortic or tricuspid regurgitation. No patient had any degree of AV block, although one developed a transient left anterior fascicular block. Holter evaluation, obtained in all patients, was unremarkable in all.

CONCLUSIONS: This early cohort experience using a novel adapted transcatheter closure device for pmVSD suggests that the procedure is feasible, safe, and effective.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23703890
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"

We want to hear from doctors like you!

Take a second to answer a survey question.