Transcatheter closure of perimembranous ventricular septal defect with the Amplatzer(®) membranous VSD occluder 2: initial world experience and one-year follow-up.
Catheterization and Cardiovascular Interventions 2014 March 2
OBJECTIVES: To describe the initial world experience and mid-term follow-up of perimembranous ventricular septal defect (pmVSD) closure with a newly designed occluder.
BACKGROUND: Transcatheter closure of pmVSDs has been associated with a substantial risk of complete heart block, prompting many centers to abandon this intervention.
METHODS: A prospective multicenter cohort study was conducted on patients with pmVSD undergoing catheter closure using the Amplatzer(®) Membranous VSD Occluder 2 in the initial 4 pilot centers.
RESULTS: Nineteen patients, median age 6 years (range 1.4-62 years), were enrolled and followed for 14 ± 3 months (range 8-20 months). The median weight was 26 kg (range 9.3-96 kg) and the mean Qp/Qs ratio was 1.8 ± 0.7. The defect on left ventricular side measured 9.9 ± 3.5 mm and the orifice on right ventricular side was 8.1 ± 2.8 mm by echocardiography. Mean device size was 9.4 ± 2.4 mm (range 5-14 mm). An eccentric device was employed in 9 patients (47%) and a concentric device in 10 (53%). Overall, 18 patients (95%) had successful device implants. Procedural time was 122 ± 39 min. There were no procedural complications. Mild residual shunting was initially observed in 14 (78%) patients. At last follow-up, mild residual shunting persisted in only 3 (17%) patients. There was no significant increase in aortic or tricuspid regurgitation. No patient had any degree of AV block, although one developed a transient left anterior fascicular block. Holter evaluation, obtained in all patients, was unremarkable in all.
CONCLUSIONS: This early cohort experience using a novel adapted transcatheter closure device for pmVSD suggests that the procedure is feasible, safe, and effective.
BACKGROUND: Transcatheter closure of pmVSDs has been associated with a substantial risk of complete heart block, prompting many centers to abandon this intervention.
METHODS: A prospective multicenter cohort study was conducted on patients with pmVSD undergoing catheter closure using the Amplatzer(®) Membranous VSD Occluder 2 in the initial 4 pilot centers.
RESULTS: Nineteen patients, median age 6 years (range 1.4-62 years), were enrolled and followed for 14 ± 3 months (range 8-20 months). The median weight was 26 kg (range 9.3-96 kg) and the mean Qp/Qs ratio was 1.8 ± 0.7. The defect on left ventricular side measured 9.9 ± 3.5 mm and the orifice on right ventricular side was 8.1 ± 2.8 mm by echocardiography. Mean device size was 9.4 ± 2.4 mm (range 5-14 mm). An eccentric device was employed in 9 patients (47%) and a concentric device in 10 (53%). Overall, 18 patients (95%) had successful device implants. Procedural time was 122 ± 39 min. There were no procedural complications. Mild residual shunting was initially observed in 14 (78%) patients. At last follow-up, mild residual shunting persisted in only 3 (17%) patients. There was no significant increase in aortic or tricuspid regurgitation. No patient had any degree of AV block, although one developed a transient left anterior fascicular block. Holter evaluation, obtained in all patients, was unremarkable in all.
CONCLUSIONS: This early cohort experience using a novel adapted transcatheter closure device for pmVSD suggests that the procedure is feasible, safe, and effective.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app