Automated 3D analysis of multislice computed tomography to define the line of perpendicularity of the aortic annulus and of the implanted valve: benefit on planning transcatheter aortic valve replacement

Mariam Samin, Francis Juthier, Camille Van Belle, Pierfrancesco Agostoni, Jolanda Kluin, Pieter R Stella, Faiez Ramjankhan, Ricardo P J Budde, Gertjan Sieswerda, Emanuela Algeri, Ahmed Elkalioubie, Anouar Belkacemi, Michel E Bertrand, Pieter A Doevendans, Eric Van Belle
Catheterization and Cardiovascular Interventions 2014 January 1, 83 (1): E119-27
23592336

AIMS: We aimed to determine whether preprocedural analysis of multislice computed tomography (MDCT) scan could accurately predict the "line of perpendicularity" (LP) of the aortic annulus and corresponding C-arm angulations required for prosthesis delivery.

METHODS AND RESULTS: A 3D analysis of preprocedural MDCT dedicated to define the LP of the aortic annulus was performed in 60 consecutive patients referred for transcatheter aortic valve replacement (TAVR). In 24 patients, the analysis was performed retrospectively to evaluate reproducibility. In 11 patients of this cohort, additional fluoroscopy and MDCT were performed postprocedure to compare the LP of the aortic annulus and the LP of the implanted bioprosthesis. In 36 patients, the analysis was performed prospectively and the results were available at the time of the procedure. In those 36 patients, the postprocedure fluoroscopy-defined LP of the implanted bioprosthesis was used to validate the LP of the aortic annulus as predicted by MDCT. Intraobserver and interobserver reproducibility of the 3D analysis of MDCT to define the LP of the aortic annulus (κ = 1 and 0.94, respectively) and of the bioprosthesis (κ= 1 and 1, respectively) were excellent. Comparison between the LP of the aortic annulus and the LP of the bioprosthesis showed that the two LPs were virtually identical, demonstrating both self-centering of the device during implantation and the possibility to use the LP of the implanted bioprosthesis as a surrogate of the LP of the aortic annulus. In the prospective cohort, the ability of MDCT analysis to predict the LP of the aortic annulus was very good (accuracy = 94% and κ = 0.89).

CONCLUSION: Automated 3D analysis of preimplantation MDCT accurately predicts the LP of the aortic annulus and the corresponding C-arm position required for TAVR.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23592336
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"