A comparison of noninvasive bioreactance with oesophageal Doppler estimation of stroke volume during open abdominal surgery: an observational study

Daniel H Conway, Osman A Hussain, Iain Gall
European Journal of Anaesthesiology 2013, 30 (8): 501-8
23549128

CONTEXT: The anaesthetist must maintain tissue perfusion by ensuring optimal perioperative fluid balance. This can be achieved using less invasive cardiac output monitors such as oesophageal Doppler monitoring (ODM). Other less invasive cardiac output monitors using bio-impedence technology (noninvasive cardiac output monitoring, NICOM) may have a role in monitoring the circulation and informing fluid management decisions.

OBJECTIVE: To compare estimates of stroke volume from ODM with those from NICOM, a noninvasive monitor using bioreactance, a modification of transthoracic bio-impedence.

DESIGN: An observational study.

SETTING: Manchester Royal Infirmary, UK. Data collected in 2011 and 2012.

PARTICIPANTS: Twenty-two patients scheduled for major, open abdominal surgery. Reasons for noninclusion: atrial fibrillation; heart failure; oesophageal disease; lack of capacity; and known sensitivity to colloid.

INTERVENTION: All patients had oesophageal Doppler cardiac output monitoring as a standard element of anaesthesia care. We placed NICOM Bioreactance electrodes and recorded stroke volume estimates from both devices. Fluid challenges were given by the anaesthetist and the haemodynamic responses were recorded.

MAIN OUTCOME MEASURE: Stroke volume during surgery. The Bland-Altman method was used to compare bias and limits of agreement for stroke volume and cardiac output. Fluid responders were defined as patients who increased stroke volume by at least 10% after fluid loading. The precision of each device was calculated during periods of haemodynamic stability.

RESULTS: We made 788 acceptable measurements of cardiac output. The bias was -6.9 ml and the limits of agreement were -22.9 to 36.8 ml. The percentage error was 57%. Average precision for both the ODM and NICOM were similar, 8.5% (SD 5.4%) and 8.7% (SD 3.2%). The concordance for the stroke volume change following fluid challenge was 90.5%. Both devices produced unacceptable readings with electrical diathermy.

CONCLUSION: Simultaneous stroke volume estimations made by noninvasive Bioreactance (NICOM) and oesophageal Doppler showed bias and limits of agreement that are not clinically acceptable. The measurements made by these two devices cannot be regarded as interchangeable.

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
23549128
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"