Comparison of four microwave ablation devices: an experimental study in ex vivo bovine liver.

Radiology 2013 July
PURPOSE: To compare volume, sphericity, and short-axis diameter of the coagulation zone of four commercially available microwave ablation systems with three technical concepts in an ex vivo setting and to formulate mathematical models to predict these quantities.

MATERIALS AND METHODS: Two high-power systems (systems A and B), a system that enables simultaneous use of three antennas (system C), and a non-perfusion-cooled system that automatically adapts power and frequency (system D) were tested in ex vivo bovine livers (108 ablations). Coagulation volume, sphericity, and mean short-axis diameter were assessed, and mathematical functions were fitted for each system and assessed with the coefficient of determination (R(2)). Analysis of variance and Tukey post hoc tests were used for interdevice comparison after 5 and 10 minutes and after maximum recommended ablation time.

RESULTS: Volume and short-axis diameter were determined by using a mathematical model for every system, with coefficients of determination of 0.75-0.98 and 0.70-0.97, respectively. Correlation for determination of sphericity was lower (R(2) = 0.01-0.68). Mean results with ablation performed according to manufacturer recommendations were as follows: Volume, sphericity, and short-axis diameter were 57.5 cm(3), 0.75, and 43.4 mm, respectively, for system A; 72.3 cm(3), 0.68, and 45.5 mm, respectively, for system B; 17.1 cm(3), 0.58, and 26.8 mm, respectively, for system C (one antenna); 76.5 cm(3), 0.89, and 50.6 mm, respectively, for system C (three antennas); and 56.0 cm(3), 0.64, and 40.9 mm, respectively, for system D. Systems A (mean volume, 52.4 cm(3) ± 4.5 [standard deviation]) and B (39.4 cm(3) ± 1.7) reach large ablation zones with 5-minute ablation.

CONCLUSION: The largest ablation zone is obtained with systems B and C (three antennas) under maximum recommended ablation duration and with system A under short ablation time. The most spherical zone is obtained with system C (three antennas).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app