Analysis of induced radionuclides in replacement parts and liquid wastes in a medical cyclotron solely used for production of 18F for [18F]FDG.
Applied Radiation and Isotopes 2013 April
Radioactivities produced in replacement parts and liquid wastes in a medical cyclotron used to produce (18)F for [(18)F]FDG with 10MeV protons were analyzed. Nineteen radionuclides were found in the replacement parts and liquid wastes. Among them, long-lived (56)Co in the Havar foils is critical in terms of radioactive waste management. The estimated dose level of exposure for the operating staff during the replacement of parts was around 310μSv/y, which is smaller than the recommended dose limit for workers.