Dosimetric influences of rotational setup errors on head and neck carcinoma intensity-modulated radiation therapy treatments

Weihua Fu, Yong Yang, Ning J Yue, Dwight E Heron, M Saiful Huq
Medical Dosimetry: Official Journal of the American Association of Medical Dosimetrists 2013, 38 (2): 125-32
23266161
The purpose of this work is to investigate the dosimetric influence of the residual rotational setup errors on head and neck carcinoma (HNC) intensity-modulated radiation therapy (IMRT) with routine 3 translational setup corrections and the adequacy of this routine correction. A total of 66 kV cone beam computed tomography (CBCT) image sets were acquired on the first day of treatment and weekly thereafter for 10 patients with HNC and were registered with the corresponding planning CT images, using 2 3-dimensional (3D) rigid registration methods. Method 1 determines the translational setup errors only, and method 2 determines 6-degree (6D) setup errors, i.e., both rotational and translational setup errors. The 6D setup errors determined by method 2 were simulated in the treatment planning system and were then corrected using the corresponding translational data determined by method 1. For each patient, dose distributions for 6 to 7 fractions with various setup uncertainties were generated, and a plan sum was created to determine the total dose distribution through an entire course and was compared with the original treatment plan. The average rotational setup errors were 0.7°± 1.0°, 0.1°±1.9°, and 0.3°±0.7° around left-right (LR), anterior-posterior (AP), and superior-inferior (SI) axes, respectively. With translational corrections determined by method 1 alone, the dose deviation could be large from fraction to fraction. For a certain fraction, the decrease in prescription dose coverage (Vp) and the dose that covers 95% of target volume (D95) could be up to 15.8% and 13.2% for planning target volume (PTV), and the decrease in Vp and the dose that covers 98% of target volume (D98) could be up to 9.8% and 5.5% for the clinical target volume (CTV). However, for the entire treatment course, for PTV, the plan sum showed that the average Vp was decreased by 4.2% and D95 was decreased by 1.2 Gy for the first phase of IMRT with a prescription dose of 50 Gy. For CTV, the plan sum showed that the average Vp was decreased by 0.8% and D98, relative to prescription dose, was not decreased. Among these 10 patients, the plan sum showed that the dose to 1-cm(3) spinal cord (D(1 cm(3))) increased no more than 1 Gy for 7 patients and more than 2 Gy for 2 patients. The average increase in D(1 cm(3)) was 1.2 Gy. The study shows that, with translational setup error correction, the overall CTV Vp has a minor decrease with a 5-mm margin from CTV to PTV. For the spinal cord, a noticeable dose increase was observed for some patients. So to decide whether the routine clinical translational setup error correction is adequate for this HNC IMRT technique, the dosimetric influence of rotational setup errors should be evaluated carefully from case to case when organs at risk are in close proximity to the target.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23266161
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"