Evaluation of TEG(®) and RoTEM(®) inter-changeability in trauma patients.

Injury 2013 May
BACKGROUND: Massive haemorrhage is a leading cause of preventable deaths in trauma. Traumatic coagulopathy is frequently present early after trauma, and is associated with increased mortality. A number of recent trials suggest that viscoelastic haemostatic assays (VHA), such as thromboelastography and thromboelastometry, are useful tools in guiding transfusion. Treatment algorithms exist for the use of VHAs but are not validated in traumatic haemorrhage. In this study we examined the inter-changeability of two commonly used VHAs, TEG(®) and RoTEM(®).

METHODS: A total of 184 trauma patients over the age of 18, requiring full trauma team activation, were included at three different hospitals in three different countries (Copenhagen, Denmark, San Francisco, CA, USA and Oslo, Norway). Blood samples were drawn immediately upon arrival, and TEG(®) and RoTEM(®) analyzed simultaneously. Correlations were calculated using. Spearman's rank correlation coefficient. Agreement was evaluated by Bland-Altman plots and calculation of limits of agreement.

RESULTS: The mean ISS in the total population was 17, and the mortality was 16.5%. Mean base excess was -2.8 (SD: 4.2). The correlation coefficient for corresponding values for the two devices was 0.24 for the R-time vs CT in all centres combined. For the K-time vs CFT the correlation was 0.48, for the α-angleTEG vs α-angleRoTEM 0.44, and for MA vs MCF 0.76. Limits of agreement exceeded the preset clinically acceptable deviation of 10% for all variables in all centres except for MA/MCF in one centre (Copenhagen). Generally, correlation coefficients were lower and agreement poorer in the one centre (Oslo) where measurements were performed bedside by clinicians.

CONCLUSION: Inter-changeability between TEG(®) and RoTEM(®) is limited in the trauma setting. Agreement seems poorer when clinicians operate the devices. Development and validation of separate treatment algorithms for the two devices is required.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app