COMPARATIVE STUDY
JOURNAL ARTICLE
MULTICENTER STUDY

Prospective comparison of valve regurgitation quantitation by cardiac magnetic resonance imaging and transthoracic echocardiography

Peter J Cawley, Christian Hamilton-Craig, David S Owens, Eric V Krieger, Wendy E Strugnell, Lee Mitsumori, Caryn L D'Jang, Rebecca G Schwaegler, Khanh Q Nguyen, Bianca Nguyen, Jeffrey H Maki, Catherine M Otto
Circulation. Cardiovascular Imaging 2013 January 1, 6 (1): 48-57
23212272

BACKGROUND: Both transthoracic echocardiography (TTE) and cardiac magnetic resonance (CMR) imaging allow quantification of chronic aortic regurgitation (AR) and mitral regurgitation (MR). We hypothesized that CMR measurement of regurgitant volume (RVol) is more reproducible than TTE.

METHODS AND RESULTS: TTE and CMR performed on the same day in 57 prospectively enrolled adults (31 with AR, 26 with MR) were measured by 2 independent physicians. TTE RVol(AR) was calculated as Doppler left ventricular outflow minus inflow stroke volume. RVol(MR) was calculated by both the proximal isovelocity surface area method and Doppler volume flow at 2 sites. CMR RVol(AR) was calculated by phase-contrast velocity mapping at the aortic sinuses and RVol(MR) as total left ventricular minus forward stroke volume. Intraobserver and interobserver variabilities were similar. For AR, the Bland-Altman mean interobserver difference in RVol was -0.7 mL (95% confidence interval [CI], -5 to 4) for CMR and -9 mL (95% CI, -53 to -36) for TTE. The Pearson correlation was higher (P=0.001) between CMR (0.99) than TTE readers (0.89). For MR, the Bland-Altman mean difference in RVol between observers was -4 mL (95% CI, -21 to 13) for CMR compared with 0.7 mL (95% CI, -30 to 32) for the proximal isovelocity surface area and -10 mL (95% CI, -76 to 56) for TTE volume flow at 2 sites. Correlation was similar for CMR (0.94) versus TTE readers (0.90 for the proximal isovelocity surface area).

CONCLUSIONS: Compared with TTE, CMR has lower intraobserver and interobserver variabilities for RVol(AR), suggesting CMR may be superior for serial measurements. Although RVol(MR) is similar by TTE and CMR, variability in measured RVol by both approaches suggests that caution is needed in clinical practice.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23212272
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"