Posterior wall capture and femoral artery stenosis following use of StarClose closing device: diagnosis and therapy.
Cardiovascular and Interventional Radiology 2013 October
A case of femoral artery obstruction following application of a StarClose type arterial puncture closing device (APCD) is presented. Ultrasonographic and angiographic imaging of this complication was obtained. The posterior wall of the vessel was accidentally caught in the anchoring element of the nitinol clip. This complication was successfully resolved by endovascular treatment and the implantation of a stent.