Systematic error in lung nodule volumetry: effect of iterative reconstruction versus filtered back projection at different CT parameters.

OBJECTIVE: Iterative reconstruction potentially can reduce radiation dose compared with filtered back projection (FBP) for chest CT. This is especially important for repeated CT scanning, as is the case in patients with indeterminate lung nodules. It is currently unknown whether absolute nodule volumes measured with iterative reconstruction are comparable to those measured with FBP. We compared nodule volumes measured with iterative reconstruction and FBP at different CT parameters.

MATERIALS AND METHODS: An anthropomorphic chest phantom was scanned using a 256-MDCT scanner at various tube voltages and tube current-time products. Raw data were reconstructed using FBP or a commercially available iterative reconstruction algorithm. Five inserted nodules with 100 HU radiodensity and different sizes (3, 5, 8, 10, and 12 mm) were measured by two observers using semiautomatic software. Volumetric nodule measurements were performed using thin-slice reconstructions.

RESULTS: For very small nodules (volume, 14.1 mm(3); diameter, 3 mm), FBP and iterative reconstruction measurements exhibited large errors and overestimated the nodule size by up to 160%. For larger nodules (volume, ≥ 65.4 mm(3); diameter, ≥ 5 mm), CT underestimated the actual size, but errors were small (within 25%) and remained small when the tube voltage and tube current-time product were reduced, even without iterative reconstruction.

CONCLUSION: In a phantom model, no clinically relevant differences beyond reported interscan variation levels between lung nodule volumes were measured in nodules 5 mm or larger at reduced tube voltage and tube current-time product, with radiation dose reductions up to 90.6% for both FBP and iterative reconstruction, suggesting that it is safe to convert FBP protocols to iterative reconstruction and reduce tube voltage and tube current-time product for lung nodule follow-up. CT appears to slightly underestimate actual nodule volume.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app