The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery.
Anesthesia and Analgesia 2012 December
BACKGROUND: The most common method for determining the hemoglobin concentration is to draw blood from a patient. However, the Radical-7 Pulse CO-Oximeter (Masimo Corporation, Irvine, CA) can noninvasively provide continuous hemoglobin concentration (SpHb). In our study we compared noninvasive measurements of SpHb with simultaneous laboratory measurements of total hemoglobin in arterial blood samples taken from children (tHb).
METHODS: Arterial blood samples were analyzed using a laboratory CO-oximeter, and SpHb was simultaneously recorded in pediatric patients undergoing neurosurgery. When patients met the criteria for hypovolemia, 10 mL/kg of colloids or red blood cells were administered over 10 minutes. SpHb and tHb data were collected before and after intravascular volume resuscitation. The relationship between SpHb and tHb was assessed using a 4-quadrant plot, linear regression, mixed-effect model, and modified Bland-Altman analyses.
RESULTS: One hundred nineteen paired samples were analyzed. The correlation coefficient between SpHb and tHb was 0.53 (P < 0.001), whereas that of change in SpHb versus change in tHb was 0.75 (P < 0.001). The average difference (bias) between tHb and SpHb was 0.90 g/dL (95% confidence interval [CI], 0.48-1.32 g/dL) and 1 standard deviation of the difference (sd) was 1.35 g/dL. The concordance rate (a measure of the number of data points that are in 1 of the 2 quadrants of agreement) determined using a 4-quadrant plot was 93%. The correlation coefficient between SpHb and tHb after intravascular volume resuscitation was 0.58 (P < 0.001), whereas that of changes in SpHb and tHb was 0.87 (P < 0.001). The bias immediately after volume resuscitation was 1.18 g/dL (95% CI, 0.81-1.55 g/dL), and sd was 1.28 g/dL with a concordance rate of 94.4%. The bias was -0.03 g/dL when tHb was ≥11 g/dL, which was significantly lower in comparison with biases when tHb <9 g/dL (1.24 g/dL) and tHb was 9-11 g/dL (1.17 g/dL) (P = 0.004).
CONCLUSION: The Radical-7 Pulse CO-Oximeter can be useful as a trend monitor in children during surgery even immediately after intravascular volume expanders are administered. However, it is advisable to confirm the baseline hemoglobin level and to consider the influence of tHb level on the bias. In addition, one should be cautious with regard to using SpHb alone when making transfusion decisions.
METHODS: Arterial blood samples were analyzed using a laboratory CO-oximeter, and SpHb was simultaneously recorded in pediatric patients undergoing neurosurgery. When patients met the criteria for hypovolemia, 10 mL/kg of colloids or red blood cells were administered over 10 minutes. SpHb and tHb data were collected before and after intravascular volume resuscitation. The relationship between SpHb and tHb was assessed using a 4-quadrant plot, linear regression, mixed-effect model, and modified Bland-Altman analyses.
RESULTS: One hundred nineteen paired samples were analyzed. The correlation coefficient between SpHb and tHb was 0.53 (P < 0.001), whereas that of change in SpHb versus change in tHb was 0.75 (P < 0.001). The average difference (bias) between tHb and SpHb was 0.90 g/dL (95% confidence interval [CI], 0.48-1.32 g/dL) and 1 standard deviation of the difference (sd) was 1.35 g/dL. The concordance rate (a measure of the number of data points that are in 1 of the 2 quadrants of agreement) determined using a 4-quadrant plot was 93%. The correlation coefficient between SpHb and tHb after intravascular volume resuscitation was 0.58 (P < 0.001), whereas that of changes in SpHb and tHb was 0.87 (P < 0.001). The bias immediately after volume resuscitation was 1.18 g/dL (95% CI, 0.81-1.55 g/dL), and sd was 1.28 g/dL with a concordance rate of 94.4%. The bias was -0.03 g/dL when tHb was ≥11 g/dL, which was significantly lower in comparison with biases when tHb <9 g/dL (1.24 g/dL) and tHb was 9-11 g/dL (1.17 g/dL) (P = 0.004).
CONCLUSION: The Radical-7 Pulse CO-Oximeter can be useful as a trend monitor in children during surgery even immediately after intravascular volume expanders are administered. However, it is advisable to confirm the baseline hemoglobin level and to consider the influence of tHb level on the bias. In addition, one should be cautious with regard to using SpHb alone when making transfusion decisions.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app