In vitro experiment for verification of the tandem shunt valve system: a novel method for treating hydrocephalus by flexibly controlling cerebrospinal fluid flow and intracranial pressure.
Journal of Neurosurgery. Pediatrics 2013 January
OBJECT: The CSF shunt valve is a medical device whose main function is to regulate intracranial pressure and drain excess CSF. The authors have developed a new therapeutic method for treating hydrocephalus, namely the tandem shunt valve system, which has the potential of flexibly controlling the CSF flow rate and intracranial pressure in patients.
METHODS: The properties of the tandem system were verified by performing in vitro experiments. An in vitro system with a manometer was built to measure pressure and flow rates of water in open systems using the Codman Hakim Programmable Valve and the Strata adjustable pressure programmable valve. A single valve and 2 single shunt valves connected in series (the tandem shunt valve system) were connected to the manometer to check the final pressure.
RESULTS: Conventional single shunt valve systems require valve pressures to be set higher to slow down the CSF flow rate, which inevitably results in a higher final pressure. On the other hand, the tandem shunt valve system uses the combination of 2 valves to slow the CSF flow rate without increasing the final pressure.
CONCLUSIONS: The authors succeeded in experimentally demonstrating in vitro results of tandem systems and their effectiveness by applying a model to show that the valve with the higher pressure setting determined the final pressure of the entire system and the flow rate became slower than single shunt valve systems.
METHODS: The properties of the tandem system were verified by performing in vitro experiments. An in vitro system with a manometer was built to measure pressure and flow rates of water in open systems using the Codman Hakim Programmable Valve and the Strata adjustable pressure programmable valve. A single valve and 2 single shunt valves connected in series (the tandem shunt valve system) were connected to the manometer to check the final pressure.
RESULTS: Conventional single shunt valve systems require valve pressures to be set higher to slow down the CSF flow rate, which inevitably results in a higher final pressure. On the other hand, the tandem shunt valve system uses the combination of 2 valves to slow the CSF flow rate without increasing the final pressure.
CONCLUSIONS: The authors succeeded in experimentally demonstrating in vitro results of tandem systems and their effectiveness by applying a model to show that the valve with the higher pressure setting determined the final pressure of the entire system and the flow rate became slower than single shunt valve systems.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app