COMPARATIVE STUDY
JOURNAL ARTICLE

Cyclic test comparison of all-inside device and inside-out sutures for radial meniscus lesion repair: an in vitro porcine model study

Yee Han Dave Lee, John Nyland, Robert Burden, David N M Caborn
Arthroscopy: the Journal of Arthroscopic & related Surgery 2012, 28 (12): 1873-81
23085200

PURPOSE: To compare biomechanical fixation and gapping characteristics of a new all-inside meniscus repair method for radial meniscus lesion repair versus conventional inside-out suture repair under submaximal cyclic loading and load-to-failure test conditions.

METHODS: Fresh-frozen porcine tibiae with attached lateral menisci and joint capsules were harvested and stored for 48 hours at -20°C. After thawing for 12 hours, equivalent-size healthy specimens were randomly assigned to 2 groups of 8 specimens each. Standardized radial lesions were repaired with the Sequent device (ConMed Linvatec, Largo, FL) (group 1) or conventional inside-out suturing with No. 2-0 braided polyester suture (group 2). Repaired specimens were placed in custom clamps and mounted on a servohydraulic device. After a 2-N preload, specimens were cycled from 5 to 20 N (0.1 Hz), before undergoing 1,000 submaximal loading cycles between 5 and 20 N (0.5 Hz). A 40-second delay at 100, 500, and 1,000 cycles enabled digital photographs to be taken for gapping measurement determination. Specimens then underwent load-to-failure testing (12.5 mm/s). Fixation failure mode was documented.

RESULTS: Group displacement did not differ after 1, 100, 500, and 1,000 submaximal loading cycles. Group peak gapping did not differ at 100, 500, and 1,000 submaximal loading cycles. Load at failure and displacement and stiffness during load-to-failure testing did not differ between groups. During load-to-failure testing, all-inside specimens failed by implant dislodgement from the meniscus periphery whereas the inside-out repaired specimens failed by suture rupture.

CONCLUSIONS: Under controlled in vitro biomechanical test conditions, the all-inside device provided radial meniscus lesion fixation that was comparable, but not superior, to conventional inside-out suturing.

CLINICAL RELEVANCE: The all-inside radial lateral meniscus lesion repair method may provide comparable fixation to conventional inside-out suturing without the need for additional incisions and their associated neurovascular injury risks.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23085200
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"