One-year outcomes of the Surgical Treatment of Aortic Stenosis With a Next Generation Surgical Aortic Valve (TRITON) trial: a prospective multicenter study of rapid-deployment aortic valve replacement with the EDWARDS INTUITY Valve System.

OBJECTIVES: A new class of rapid-deployment aortic valves has emerged with the potential to simplify minimally invasive aortic surgery and reduce crossclamp and cardiopulmonary bypass times. We report the 1-year clinical outcomes of aortic valve replacement with the EDWARDS INTUITY Valve System (Edwards Lifesciences LLC, Irvine, Calif) in the Surgical Treatment of Aortic Stenosis With a Next Generation Surgical Aortic Valve (TRITON) trial.

METHODS: Seventeen surgeons from 6 European centers treated 152 consecutive patients with aortic stenosis requiring valve replacement in a prospective, single-arm trial. A stented trileaflet bovine pericardial bioprosthesis with a balloon-expandable, cloth-covered stent frame at the inflow aspect was implanted in 146 patients (mean age, 75.5 ± 6.7 years; 52.7% were female). Five valve sizes were evaluated (19-27 mm); 58.9% of cases had isolated aortic valve replacement, and 41.1% of cases involved concomitant procedures. Minimally invasive surgical approaches occurred in 48.8% of the isolated aortic valve replacements. Patients were followed at discharge, 3 months, and 1 year postoperatively.

RESULTS: Implantation success was 96.1% (146/152), early valve-related mortality was 1.4% (2/146), and cumulative survival was 92.5% at a mean follow-up of 9.8 ± 5.1 months. Crossclamp time for isolated aortic valve replacement was 41.1 ± 10.6 minutes. Independent core laboratory-adjudicated mean effective orifice area and aortic valve pressure gradient were 1.7 ± 0.2 cm(2) and 8.8 ± 3.0 mm Hg at 3 months, and 1.7 ± 0.2 cm(2) and 8.4 ± 3.4 mm Hg at 1 year, respectively.

CONCLUSIONS: Implantation of the EDWARDS INTUITY Valve System is feasible, safe, and efficacious for aortic valve replacement. Aortic crossclamp and cardiopulmonary bypass times were reduced compared with those for conventional aortic valve replacement. Early hemodynamic performance was excellent and remained so up to 1 year.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app