JOURNAL ARTICLE

Adaptive off-line protocol for prostate external radiotherapy with cone beam computer tomography

M Piziorska, P Kukołowicz, A Zawadzka, M Pilichowska, P Pęczkowski
Strahlentherapie und Onkologie: Organ der Deutschen Röntgengesellschaft ... [et Al] 2012, 188 (11): 1003-9
23053159

PURPOSE: The goal of this work was to prepare and to evaluate an off-line adaptive protocol for prostate teleradiotherapy with kilovoltage cone beam computer tomography (CBCT).

PATIENTS AND METHODS: Ten patients with localized prostate carcinoma treated with external beams underwent image-guided radiotherapy. In total, 162 CBCT images were collected. Position of prostate and pubis symphysis (PS) with respect to the isocenter were measured off-line. Using the CBCT scans obtained in the first three fractions the average position of prostate in relation (AvPosPr) to PB was calculated. On each CBCT scan, the position of prostate with respect to AvPosPr was calculated and cumulative histogram of prostate displacement with respect to AvPosPr was prepared. Using this data, the adaptive protocol was prepared in which (1) based on the CBCT made in the first three fractions the AvPosPr to PS is obtained, (2) in all other fractions two orthogonal images are acquired and if for any direction set-up error exceeds 0.2 cm the patient's position is corrected, and (3) additionally, the patient's position is corrected if the AvPosPr exceeds 0.2 cm in any direction. To evaluate the adaptive protocol for 30 consecutive patients, the CBCT was also made in 10th and 21st fraction.

RESULTS: For the first 10 patients, the results revealed that the prostate was displaced in relation to AvPosPr >0.7 cm in the vertical and longitudinal directions only on 4 and 5 images of 162 CBCT images, respectively. For the lateral direction, this displacement was >0.3 cm in one case. For the group of 30 patients, displacement was never >0.7, and 0.3 cm for the vertical and lateral directions. In two cases, displacements were >0.7 cm for the longitudinal direction.

CONCLUSION: Implementation of the proposed adaptive procedure based on the on-line set-up error elimination followed by a reduction of systematic internal error enables reducing the CTV-PTV margin to 0.7, 0.7, and 0.4 cm for the vertical, longitudinal, and lateral directions, respectively.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
23053159
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"