Emphysema quantification by low-dose CT: potential impact of adaptive iterative dose reduction using 3D processing.

OBJECTIVE: The purpose of this study is to investigate the effect of a novel reconstruction algorithm, adaptive iterative dose reduction using 3D processing, on emphysema quantification by low-dose CT.

MATERIALS AND METHODS: Twenty-six patients who had undergone standard-dose (150 mAs) and low-dose (25 mAs) CT scans were included in this retrospective study. Emphysema was quantified by several quantitative measures, including emphysema index given by the percentage of lung region with low attenuation (lower than -950 HU), the 15th percentile of lung density, and size distribution of low-attenuation lung regions, on standard-dose CT images reconstructed without adaptive iterative dose reduction using 3D processing and on low-dose CT images reconstructed both without and with adaptive iterative dose reduction using 3D processing. The Bland-Altman analysis was used to assess whether the agreement between emphysema quantifications on low-dose CT and on standard-dose CT was improved by the use of adaptive iterative dose reduction using 3D processing.

RESULTS: For the emphysema index, the mean differences between measurements on low-dose CT and on standard-dose CT were 1.98% without and -0.946% with the use of adaptive iterative dose reduction using 3D processing. For 15th percentile of lung density, the mean differences without and with adaptive iterative dose reduction using 3D processing were -6.67 and 1.28 HU, respectively. For the size distribution of low-attenuation lung regions, the ranges of the mean relative differences without and with adaptive iterative dose reduction using 3D processing were 21.4-85.5% and -14.1% to 11.2%, respectively. For 15th percentile of lung density and the size distribution of low-attenuation lung regions, the agreement was thus improved by the use of adaptive iterative dose reduction using 3D processing.

CONCLUSION: The use of adaptive iterative dose reduction using 3D processing resulted in greater consistency of emphysema quantification by low-dose CT, with quantification by standard-dose CT.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app