Prosthesis oversizing in balloon-expandable transcatheter aortic valve implantation is associated with contained rupture of the aortic root.
Circulation. Cardiovascular Interventions 2012 August 2
BACKGROUND: To retrospectively investigate the potential cause of contained rupture of the aortic root in balloon-expandable transcatheter aortic valve implantation (TAVI) by means of pre- and postinterventional multislice computed tomography.
METHODS AND RESULTS: Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm(2)) underwent balloon-expandable TAVI using the EdwardsSAPIEN Transcatheter Heart Valve (23 mm, n=19; 26 mm, n=50; 29 mm, n=3). Aortic annulus dimensions were quantified by multislice computed tomography-based cross-sectional area assessment and average diameter calculation (CAAD) before and after TAVI. Post-TAVI multislice computed tomography data sets were available in 65 patients; contained aortic root rupture was diagnosed in 3 patients. Pre-TAVI CAAD was 23.1±1.8 mm; post-TAVI CAAD was 22.9±1.3 mm. Median relative change in CAAD pre- and post-TAVI was -0.5% (interquartile range, 3.6%). Relative increase of 5% to 10% was observed in 4 patients (1 with contained rupture), relative increase >10% in 2 patients, both with contained rupture. Mean relative oversizing, calculated as the relative difference in diameter between pre-TAVI CAAD and nominal diameter of the selected prosthesis, was 9.8%±7.8%. Relative oversizing was significantly higher in patients with contained rupture compared with patients without contained rupture (24.6%±5.4% versus 9.1%±6.6%; P<0.001). Relative oversizing ≥20% occurred in 6 patients (3 with contained rupture).
CONCLUSIONS: Contained rupture of the aortic root in balloon-expandable TAVI is associated with severe prosthesis oversizing. Multislice computed tomography-based assessment of aortic annulus dimension in conjunction with adapted sizing guidelines may reduce the incidence of severe oversizing.
METHODS AND RESULTS: Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm(2)) underwent balloon-expandable TAVI using the EdwardsSAPIEN Transcatheter Heart Valve (23 mm, n=19; 26 mm, n=50; 29 mm, n=3). Aortic annulus dimensions were quantified by multislice computed tomography-based cross-sectional area assessment and average diameter calculation (CAAD) before and after TAVI. Post-TAVI multislice computed tomography data sets were available in 65 patients; contained aortic root rupture was diagnosed in 3 patients. Pre-TAVI CAAD was 23.1±1.8 mm; post-TAVI CAAD was 22.9±1.3 mm. Median relative change in CAAD pre- and post-TAVI was -0.5% (interquartile range, 3.6%). Relative increase of 5% to 10% was observed in 4 patients (1 with contained rupture), relative increase >10% in 2 patients, both with contained rupture. Mean relative oversizing, calculated as the relative difference in diameter between pre-TAVI CAAD and nominal diameter of the selected prosthesis, was 9.8%±7.8%. Relative oversizing was significantly higher in patients with contained rupture compared with patients without contained rupture (24.6%±5.4% versus 9.1%±6.6%; P<0.001). Relative oversizing ≥20% occurred in 6 patients (3 with contained rupture).
CONCLUSIONS: Contained rupture of the aortic root in balloon-expandable TAVI is associated with severe prosthesis oversizing. Multislice computed tomography-based assessment of aortic annulus dimension in conjunction with adapted sizing guidelines may reduce the incidence of severe oversizing.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app