ENGLISH ABSTRACT
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

[Evaluation of accuracy and influence factors of bedside blood glucose monitoring in critically ill patients].

OBJECTIVE: To evaluate the accuracy and influence factors of point-of-care testing (POCT) for glucose in critically ill patients.

METHODS: Two hundred and forty critically ill patients aged ≥18 years in department of critical care medicine were enrolled. According to blood glucose level (BGL) during glucose control, patients were divided into three groups: (1) hypoglycemia group, BGL<4.5 mmol/L, n=32; (2) euglycemia group, BGL 4.5-8.3 mmol/L, n=138; (3) hyperglycemia group, BGL>8.3 mmol/L, n=70. The blood samples from vein, artery and capillary of patients were collected synchronically and the blood glucose of POCT were determined with glucose oxidase (GOD) and glucose dehydrogenase (GDH) methods, respectively, compared with blood glucose reference values of laboratory [hexokinase method (HK method)]. The accuracy of POCT for glucose and influence factors were analyzed statistically by the logistic regression method.

RESULTS: (1) The inaccurate rates of glucose values in blood samples from vein, artery and capillary in hypoglycemia group (GDH method: 25.00%, 40.62%, 40.62%; GOD method: 59.38%, 71.88%, 71.88%) were significantly higher than those in euglycemia group (GDH method: 2.90%, 9.42%, 7.97%; GOD method: 18.12%, 27.54%, 27.54%) and hyperglycemia group (GDH method: 1.43%, 8.57%, 4.28%; GOD method: 11.43%, 8.57%, 11.43%, all P<0.01). (2) The average levels of difference for the glucose reference value of laboratory and the glucose value measured by glucometry in hypoglycemia group were 0.41-0.69 mmol/L (GDH method) and 0.92-1.18 mmol/L (GOD method), in euglycemia 0.16-0.33 mmol/L and 0.77-0.90 mmol/L, in hyperglycemia group -0.06-0.18 mmol/L and 0.56-0.76 mmol/L, respectively. (3) The correlation coefficients between the laboratory and glucometry in hypoglycemia group were respectively 0.812-0.853 (GDH method) and 0.723-0.816 (GOD method). The correlation coefficients in euglycemia group were 0.862-0.890 and 0.768-0.857. They were elevated to 0.922-0.957 and 0.896-0.922 in hyperglycemia group (all P<0.01). (4)On the influence factors of accuracy in POCT for glucose, BGL, acute physiology and chronic health evaluation II (APACHEII) score and hematocrit level were into the logistic regression equation.

CONCLUSIONS: The accuracy of POCT for glucose was markedly affected by the BGL, severity of patients and hematocrit level during glucose control. Under hypoglycemia, the agreement between laboratory and glucometry measurements was obviously fallen and it tended to overestimate the patient's real glucose value.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app