Helical tomotherapy for SIB and hypo-fractionated treatments in lung carcinomas: a 4D Monte Carlo treatment planning study.

PURPOSE: To evaluate the impact of intra-fraction motion induced by regular breathing on treatment quality for helical tomotherapy treatments.

MATERIAL AND METHODS: Four patients treated by simultaneous-integrated boost (SIB) and three by hypo-fractionated stereotactic treatments (hypo-fractionated, 18 Gy/fraction) were included. All patients were coached to ensure regular breathing. For the SIB group, the tumor volume was delineated using CT information only (CTV(CT)) and the boost region was based on PET information (GTV(PET), no CTV extension). In the hypo-fractionated group, a GTV based on CT information was contoured. In both groups, ITVs were defined according to 4D data. The PTV included the ITV plus a setup error margin. The treatment was planned using the tomotherapy TPS on 3D CT images. In order to verify the impact of intra-fraction motion and interplay effects, dose calculations were performed using a previously validated Monte Carlo model of tomotherapy (TomoPen): first on the planning 3D CT ("planned dose") and second, on the 10 phases of the 4D scan. For the latter, two dose distributions, termed "interplay simulated" or "no interplay" were computed with and without beamlet-phase correlation over the 10 phases and combined using deformable dose registration.

RESULTS: In all cases, DVHs of "interplay simulated" dose distributions complied within 1% of the original clinical objectives used for planning, defined according to ICRU (report 83) and RTOG (trials 0236 and 0618) recommendations, for SIB and hypo-fractionated groups, respectively. For one patient in the hypo-fractionated group, D(mean) to the CTV(CT) was 2.6% and 2.5% higher than "planned" for "interplay simulated" and "no interplay", respectively.

CONCLUSION: For the patients included in this study, assuming regular breathing, the results showed that interplay of breathing and tomotherapy delivery motions did not affect significantly plan delivery accuracy. Hence, accounting for intra-fraction motion through the definition of an ITV volume was sufficient to ensure tumor coverage.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app