BACKGROUND: The Masimo Radical 7 (Masimo Corp., Irvine, CA, USA) pulse co-oximeter(®) calculates haemoglobin concentration (SpHb) non-invasively using transcutaneous spectrophotometry. We compared SpHb with invasive satellite-lab haemoglobin monitoring (Hb(satlab)) during major hepatic resections both under steady-state conditions and in a dynamic phase with fluid administration of crystalloid and colloid solutions.

METHODS: Thirty patients undergoing major hepatic resection were included and randomized to receive a fluid bolus of 15 ml kg(-1) colloid (n=15) or crystalloid (n=15) solution over 30 min. SpHb was continuously measured on the index finger, and venous blood samples were analysed in both the steady-state phase (from induction until completion of parenchymal transection) and the dynamic phase (during fluid bolus).

RESULTS: Correlation was significant between SpHb and Hb(satlab) (R(2)=0.50, n=543). The modified Bland-Altman analysis for repeated measurements showed a bias (precision) of -0.27 (1.06) and -0.02 (1.07) g dl(-1) for the steady-state and dynamic phases, respectively. SpHb accuracy increased when Hb(satlab) was <10 g dl(-1), with a bias (precision) of 0.41 (0.47) vs -0.26 (1.12) g dl(-1) for values >10 g dl(-1), but accuracy decreased after colloid administration (R(2)=0.25).

CONCLUSIONS: SpHb correlated moderately with Hb(satlab) with a slight underestimation in both phases in patients undergoing major hepatic resection. Accuracy increased for lower Hb(satlab) values but decreased in the presence of colloid solution. Further improvements are necessary to improve device accuracy under these conditions, so that SpHb might become a sensitive screening device for clinically significant anaemia.

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
22822043
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"