False positive rate of carbon monoxide saturation by pulse oximetry of emergency department patients.

Respiratory Care 2013 Februrary
BACKGROUND: Symptoms of carbon monoxide (CO) poisoning are non-specific. Diagnosis requires suspicion of exposure, confirmed by measuring ambient CO levels or carboxyhemoglobin (COHb). An FDA-approved pulse oximeter (Rad-57) can measure CO saturation (S(pCO)). The device accuracy has implications for clinical decision-making.

METHODS: From April 1 to August 15, 2008, study personnel measured S(pCO) and documented demographic factors at time of clinical blood draw, in a convenience sample of 1,363 subjects presenting to the emergency department at Intermountain Medical Center, Murray, Utah. The technician then assayed COHb. COHb and S(pCO) values were compared by subject; false positive or negative values were defined as S(pCO) at least 3 percentage points greater or less than COHb level, reported by the manufacturer to be ± 1 SD in performance.

RESULTS: In 1,363 subjects, 613 (45%) were male, 1,141 (84%) were light-skinned, 14 in shock, 4 with CO poisoning, and 122 (9%) met the criteria for a false positive value (range 3-19 percentage points), while 247 (18%) met the criteria for a false negative value (-13 to -3 percentage points). Risks for a false positive S(pCO) reading included being female and having a lower perfusion index. Methemoglobin, body temperature, and blood pressure also appear to influence the S(pCO) accuracy. There was variability among monitors, possibly related to technician technique, as rotation of monitors among technicians was not enforced.

CONCLUSIONS: While the Rad-57 pulse oximeter functioned within the manufacturer's specifications, clinicians using the Rad-57 should expect some S(pCO) readings to be significantly higher or lower than COHb measurements, and should not use S(pCO) to direct triage or patient management. An elevated S(pCO) could broaden the diagnosis of CO poisoning in patients with non-specific symptoms. However, a negative S(pCO) level in patients suspected of having CO poisoning should never rule out CO poisoning, and should always be confirmed by COHb.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app