JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Accuracy of intraoperative fluoroscopy with and without laser guidance in foot and ankle surgery.

BACKGROUND: The use of intraoperative fluoroscopy in orthopedic surgery involves frequent exposure to ionizing radiation in the operating room. Although a number of studies have found radiation exposure to orthopedic surgeons to be well below recommended doses, the long-term effects of low-dose radiation are not well known. Thus, all effective practical methods should be undertaken to reduce the exposure to radiation. Our purpose was to determine whether the use of a laser-aiming device improves the accuracy of intraoperative fluoroscopy to reduce, by implication, radiation exposure in the operating room.

METHODS: From March to October 2004, there were 92 consecutive cases requiring use of fluoroscopy in the orthopedic foot and ankle service of the authors' institution. The number of accurate and inaccurate images with or without the presence of a radiology technician and a laser-aiming device were compared.

RESULTS: The accuracy of imaging with the laser-aiming device was higher than the imaging without the device (p < 0.001). The accuracy of the images obtained by the surgeon was higher than the technicians' images when laser guidance was used (p= 0.027). There was no significant difference between the images obtained by the surgeon or the technicians when the aiming device was not used (p = 0.09).

CONCLUSION: The use of a laser-aiming device to help position during fluoroscopy is recommended in an effort to reduce radiation exposure in foot and ankle surgery.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app