Vena cava filter practices of a regional vascular surgery society.

BACKGROUND: Vena cava filter (VCF) use in the United States has increased dramatically with prophylactic indications for placement and the availability of low-profile retrievable devices, which are overtaking the filter market. We surveyed the practice patterns of a large group of vascular surgeons from a regional vascular surgery society to see whether they mirrored current national trends.

METHODS: A 17-question online VCF survey was offered to all members of the Southern Association of Vascular Surgery. The responses were analyzed using the χ(2) goodness of fit tests.

RESULTS: Of the 276 members surveyed, 126 (46%) responded, with 118 (93%) indicating that they placed filters during their practice. Highly significant differences were identified with each question (at least P < 0.002). Regarding the inferior vena cava, the preferred permanent filters were the Greenfield (31%), the TrapEase (15%), the Vena Tech (5%), and a variety of retrievable devices (49%). Fifty percent of the respondents placed retrievable filters selectively; 26% always placed them; and 24% never did. Filters were placed for prophylactic indications <50% of the time by 63% of the respondents. Overall, retrievable filters (when not used as permanent filters) were removed <25% of the time by 64% of the respondents and <50% of the time by 78% of the respondents. The femoral vein was the preferred access site for 84% of the respondents. Major complications were few but included filter migration to the atrium (one), atrial perforation (one), abdominal pain requiring surgical filter removal (two), inferior vena cava thrombosis (12 vena cava thrombosis--4 due to TrapEase filters), strut fracture with embolization to heart or lungs (three Bard retrievable filters), and severe tilting precluding percutaneous retrieval and protection from pulmonary emboli (8 filters with severe tilt--7 of which were Bard). Of the respondents, 59% had never placed a superior vena cava filter, and 28% had placed five or fewer.

CONCLUSIONS: Although VCF insertion overall appears safe, some complications are specific to biconical and certain retrievable filters. Given the low removal rate and lack of long-term experience with retrievable filters, routine use of these devices as permanent filters should be questioned. If used on a temporary basis, there should be a plan for filter removal at the time of implantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app