Factors predicting and having an impact on the need for a permanent pacemaker after CoreValve prosthesis implantation using the new Accutrak delivery catheter system.

OBJECTIVES: The purpose of this study was to evaluate the need for a permanent pacemaker after transcatheter aortic valve implantation with the CoreValve prosthesis (Medtronic, Inc., Minneapolis, Minnesota) using the new Accutrak delivery system (Medtronic, Inc.).

BACKGROUND: The need for a permanent pacemaker is a recognized complication after transcatheter aortic valve implantation with the CoreValve prosthesis.

METHODS: Between April 23, 2008 and May 31, 2011, 195 consecutive patients with symptomatic aortic valve stenosis underwent transcatheter aortic valve implantation using the self-expanding CoreValve prosthesis. In 124 patients, the traditional delivery system was used, and in 71 patients, the Accutrak delivery system was used.

RESULTS: There were no significant differences in baseline electrocardiographic characteristics between the traditional system and the Accutrak patients: PR interval: 153 ± 46 mm versus 165 ± 30 mm, p = 0.12; left bundle branch block: 22 (20.2%) versus 8 (12.7%), p = 0.21; right bundle branch block: 21 (19.3%) versus 8 (12.7%), p = 0.26. The depth of the prosthesis in the left ventricular outflow tract was greater with the traditional system than with the Accutrak system (9.6 ± 3.2 mm vs. 6.4 ± 3 mm, p < 0.001) and the need for a permanent pacemaker was higher with traditional system than with Accutrak (35.1% vs. 14.3%, p = 0.003). The predictors of the need for a pacemaker were the depth of the prosthesis in the left ventricular outflow tract (hazard ratio [HR]: 1.2, 95% confidence interval [CI]: 1.08 to 1.34, p < 0.001), pre-existing right bundle branch block (HR: 3.5, 95% CI: 1.68 to 7.29, p = 0.001), and use of the traditional system (HR: 27, 95% CI: 2.81 to 257, p = 0.004).

CONCLUSIONS: The new Accutrak delivery system was associated with less deep prosthesis implantation in the left ventricular outflow tract, which could be related to the lower rate of permanent pacemaker requirement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app