BACKGROUND AND PURPOSE: The Near Needle Holder (NNH) (Near Manufacturing, Camrose, Alberta, Canada) is a reusable tool to introduce a standard hollow needle for pediatric intraosseous (IO) infusion. We compared the NNH to the Cook Dieckmann (Cook Critical Care, Bloomington, IN) manual IO needle in a simulation setting.
METHODS: Study subjects were 32 physicians, nurses, and medical students participating in a trauma course in Guyana. After watching a training video and practicing under supervision, subjects were observed inserting each device into a pediatric leg model using a randomized crossover design. Outcome measures were time to successful insertion, technical complications, ease of use, and safety of each device.
RESULTS: The mean time for IO insertion (32 ± 13 seconds) was similar for both devices (P = .92). Subjects rated the NNH device equivalent in ease of use to the Cook IO needle but slightly lower in perceived safety to the user.
CONCLUSIONS: After training, all subjects successfully inserted the NNH IO device in a simulation environment, and most rated it as easy to use and safe. The NNH is a significant advance because IO needles are often not available in emergency departments in developing countries. Further studies are needed to evaluate clinical effectiveness of the NNH.
METHODS: Study subjects were 32 physicians, nurses, and medical students participating in a trauma course in Guyana. After watching a training video and practicing under supervision, subjects were observed inserting each device into a pediatric leg model using a randomized crossover design. Outcome measures were time to successful insertion, technical complications, ease of use, and safety of each device.
RESULTS: The mean time for IO insertion (32 ± 13 seconds) was similar for both devices (P = .92). Subjects rated the NNH device equivalent in ease of use to the Cook IO needle but slightly lower in perceived safety to the user.
CONCLUSIONS: After training, all subjects successfully inserted the NNH IO device in a simulation environment, and most rated it as easy to use and safe. The NNH is a significant advance because IO needles are often not available in emergency departments in developing countries. Further studies are needed to evaluate clinical effectiveness of the NNH.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app