OBJECTIVE: To report a multi-center experience with the novel Hemodialysis Reliable Outflow (HeRO) vascular access graft.

MATERIALS AND METHODS: Four centers conducted a retrospective review of end stage renal disease patients who received the HeRO device from implant to last available follow-up. Data is available on 164 patients with an accumulated 2092.1 HeRO implant months.

RESULTS: At 6 months, HeRO primary and secondary patency is 60% and 90.8%, respectively and at 12 months, 48.8% and 90.8%, respectively. At 24 months, HeRO had a primary patency of 42.9% and secondary patency was 86.7%. Interventions to maintain or re-establish patency have been required in 71.3% of patients (117/164) resulting in an intervention rate of 1.5/year. Access related infections have been reported in 4.3% patients resulting in a rate of 0.14/1000 implant days.

CONCLUSIONS: In our experience the HeRO device has performed comparably to standard AVGs and has proven superior to TDCs in terms of patency, intervention, and infection rates when compared to the peer-reviewed literature. As an alternative to catheter dependence as a means for hemodialysis access, this graft could reduce the morbidity and mortality associated with TDCs and have a profound impact on the costs associated with catheter related infections and interventions.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
22580402
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"