Intrafraction verification of gated RapidArc by using beam-level kilovoltage X-ray images.

PURPOSE: To verify the geometric accuracy of gated RapidArc treatment using kV images acquired during dose delivery.

METHODS AND MATERIALS: Twenty patients were treated using the gated RapidArc technique with a Varian TrueBeam STx linear accelerator. One to 7 metallic fiducial markers were implanted inside or near the tumor target before treatment simulation. For patient setup and treatment verification purposes, the internal target volume (ITV) was created, corresponding to each implanted marker. The gating signal was generated from the Real-time Position Management (RPM) system. At the beginning of each fraction, individualized respiratory gating amplitude thresholds were set based on fluoroscopic image guidance. During the treatment, we acquired kV images immediately before MV beam-on at every breathing cycle, using the on-board imaging system. After the treatment, all implanted markers were detected, and their 3-dimensional (3D) positions in the patient were estimated using software developed in-house. The distance from the marker to the corresponding ITV was calculated for each patient by averaging over all markers and all fractions.

RESULTS: The average 3D distance between the markers and their ITVs was 0.8 ± 0.5 mm (range, 0-1.7 mm) and was 2.1 ± 1.2 mm at the 95th percentile (range, 0-3.8 mm). On average, a left-right margin of 0.6 mm, an anterior-posterior margin of 0.8 mm, and a superior-inferior margin of 1.5 mm is required to account for 95% of the intrafraction uncertainty in RPM-based RapidArc gating.

CONCLUSION: To our knowledge, this is the first clinical report of intrafraction verification of respiration-gated RapidArc treatment in stereotactic ablative radiation therapy. For some patients, the markers deviated significantly from the ITV by more than 2 mm at the beginning of the MV beam-on. This emphasizes the need for gating techniques with beam-on/-off controlled directly by the actual position of the tumor target instead of external surrogates such as RPM.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app