We have located links that may give you full text access.
ENGLISH ABSTRACT
EVALUATION STUDIES
JOURNAL ARTICLE
[Evaluation of a portable device based on peripheral arterial tone in the detection of obstructive sleep apnea].
Zhonghua Er Bi Yan Hou Tou Jing Wai Ke za Zhi = Chinese Journal of Otorhinolaryngology Head and Neck Surgery 2012 Februrary
OBJECTIVE: To assess the accuracy of a wrist-worn device (Watch-PAT 200) in the diagnosis of obstructive sleep apnea syndrome (OSAHS).
METHODS: Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch-PAT 200 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.
RESULTS: The mean age of the subjects was (42.2 ± 12.2) years (x(-) ± s), and mean body mass index was (28.0 ± 3.9) kg/m(2). Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour, and mean PAT-AHI was (36.0 ± 29.2) events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r = 0.931, P < 0.01). A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200. At lower levels of AHI, PAT tended to overestimate disease severity, while at higher levels of AHI, Watch-PAT underestimated severity. To assess sensitivity and specificity of Watch-PAT, constructed receiver operator characteristic curves using a variety of AHI threshold values (5, 15 and 30 events per hour). For AHI ≥ 5 events per hour as threshold value, the Watch-PAT diagnosing rate was 93%, and sensitivity as well as specificity were 94.7% and 80.0%. The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%. Optimal combinations of sensitivity and specificity for the AHI threshold values (15 and 30 events per hour) were 82.6% and 100.0%, 95.0% and 95.7% respectively.
CONCLUSION: The Watch-PAT 200 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAHS, with minimal patient discomfort.
METHODS: Forty-three adult subjects with suspected OSAHS simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch-PAT 200 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. The PSG recordings were blindly manually analyzed, while Watch-PAT data were scored automatically based on the algorithm developed previously.
RESULTS: The mean age of the subjects was (42.2 ± 12.2) years (x(-) ± s), and mean body mass index was (28.0 ± 3.9) kg/m(2). Mean PSG apnea hypopnea index (AHI) was (34.9 ± 29.9) events per hour, and mean PAT-AHI was (36.0 ± 29.2) events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r = 0.931, P < 0.01). A Bland-Altman plot of PAT AHI and PSG AHI was also used to assess the accuracy of Watch-PAT 200. At lower levels of AHI, PAT tended to overestimate disease severity, while at higher levels of AHI, Watch-PAT underestimated severity. To assess sensitivity and specificity of Watch-PAT, constructed receiver operator characteristic curves using a variety of AHI threshold values (5, 15 and 30 events per hour). For AHI ≥ 5 events per hour as threshold value, the Watch-PAT diagnosing rate was 93%, and sensitivity as well as specificity were 94.7% and 80.0%. The misdiagnosis rate and missed diagnosis rate were 20.0% and 5.3%. Optimal combinations of sensitivity and specificity for the AHI threshold values (15 and 30 events per hour) were 82.6% and 100.0%, 95.0% and 95.7% respectively.
CONCLUSION: The Watch-PAT 200 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAHS, with minimal patient discomfort.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app