JOURNAL ARTICLE

Set-up errors and planning target volume margins in head and neck cancer radiotherapy: a clinical study of image guidance with on-line cone-beam computed tomography

Francesco Dionisi, Mauro Filippo Palazzi, Francesco Bracco, Maria Grazia Brambilla, Claudia Carbonini, Diego Dario Asnaghi, Angelo Filippo Monti, Alberto Torresin
International Journal of Clinical Oncology 2013, 18 (3): 418-27
22389140

BACKGROUND: Set-up errors represent a source of uncertainty in head and neck (H&N) cancer radiotherapy. The present study evaluated set-up accuracy with the use of cone-beam computed tomography (CBCT) in order to establish the proper clinical target volume (CTV) to planning target volume (PTV) margins to be adopted.

METHODS: Local set-up accuracy was analysed for 44 H&N cancer patients since the implementation of CBCT. An on-line correction protocol was adopted, with the first 3 scans used to correct systematic errors with a 3-mm action level. The overall mean displacement (M), the population systematic (Σ) and random (σ) errors and the 3D vector length were calculated. PTV margins were calculated according to the van Herk formula (2.5Σ + 0.7σ).

RESULTS: A total of 420 CBCT scans were analysed. A systematic correction was needed in 43% of patients. The value of M was <1 mm in all directions; the values of Σ and σ ranged over 1-1.2 and 1.4-1.9 mm, respectively. Pre-correction PTV margins were 3.48, 4.08 and 4.33 mm along the 3 axes. The PTV margins calculated after online correction were <2.5 mm in all directions.

CONCLUSIONS: Kilovoltage CBCT is effective in evaluating set-up accuracy in H&N patients. CTV-PTV margins of 5 mm are safe and are currently adopted at our centre; however, some special situations, such as re-irradiation or the close proximity of organs at risk and high-dose regions, could benefit from daily image registration and lower (i.e., 3 mm) margins.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
22389140
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"