COMPARATIVE STUDY
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

A new imaging method for assessment of aortic dissection using four-dimensional phase contrast magnetic resonance imaging.

INTRODUCTION: Medical management of type B aortic dissection can result in progressive dilation of the false lumen and poor long-term outcome. Recent studies using models of aortic dissection have suggested flow characteristics, such as stroke volume, velocity, and helicity, are related to aortic expansion. The aim of this study was to assess whether four-dimensional phase-contrast magnetic resonance imaging (4D PC-MRI) can accurately visualize and quantify flow characteristics in patients with aortic dissection and whether these features are related to the rate of aortic expansion.

METHODS: Twelve consecutive patients with medically treated type B thoracic aortic dissection underwent a three-dimensional (3D) MRI anatomy scan using a blood pool contrast agent. Two-dimensional phase contrast MRI data (2D PC-MRI) were acquired in the ascending and descending aorta and 4D PC-MRI data were acquired in the entire thoracic aorta. The 2D PC-MRI measurements were used to assess the quality of the 4D PC-MRI velocity data. Stroke volume, velocity, and the direction of flow were calculated using 4D PC-MRI and related to the rate of aortic expansion measured on contrast-enhanced computed tomography.

RESULTS: Comparison of 2D PC-MRI and 4D PC-MRI measurements showed good correlation (Pearson R(2) = 0.98; 95% confidence interval [CI], 0.9818-0.9953; P < .0001) and no proportional bias (bias = 1.0 mL; standard deviation, 4.6). The median aortic growth rate was 6.1 mm/y (interquartile range [IQR], 1.1-15.1 mm/y), and this correlated well with the growth rate of the false lumen (Spearman ρ = 0.62; 95% CI, 0.06-0.89; P = .0347). False lumen thrombosis (FLT) was seen in 7 of 12 patients and was not associated with reduced aortic expansion rate (FLT present: 11.4 mm/y; IQR, 3.6-21.4) vs FLT absent: 9.9 mm/y; IQR, 3.4-24.2; Mann-Whitney P = .8763). False lumen stroke volume and velocity were associated with more rapid aortic expansion (ρ = 0.80 [95% CI, 0.39-0.94; P = .0029] and ρ = 0.59 [95% CI, 0.09-0.87; P = .0480] respectively). The position of the dominant entry tear was associated with rapid expansion, which tended to be higher with distal vs proximal entry tears (distal, 21.4 mm/y [IQR, 11.4-48.9] vs proximal, 5.5 mm/y [IQR, 3.4-16.6]; Mann-Whitney P = .096). Helical flow was seen in the false lumen in 8 of 12 patients and was related to the rate of aortic expansion (ρ = 0.83, P = .0154).

CONCLUSIONS: 4D PC-MRI can be accurately applied to visualize and quantify flow characteristics in patients with aortic dissection. Stroke volume, velocity, distal dominant entry tears, and helical flow are related to the rate of aortic expansion. This study demonstrates the potential of this new imaging method. A larger prospective study is now required to measure flow characteristics and determine their predictive value for risk stratification of patients with aortic dissection.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app