Biomechanical investigation of a novel integrated device for intra-articular stabilization of the C1-C2 (atlantoaxial) joint

Peter A Robertson, Parmenion P Tsitsopoulos, Leonard I Voronov, Robert M Havey, Avinash G Patwardhan
Spine Journal: Official Journal of the North American Spine Society 2012, 12 (2): 136-42
22341395

BACKGROUND CONTEXT: The anatomy of the atlantoaxial joint makes stabilization at this level challenging. Current techniques that use transarticular screw fixation (Magerl) or segmental screw fixation (Harms) give dramatically improved stability but risk damage to the vertebral artery. A novel integrated device was designed and developed to obtain intra-articular stabilization via primary interference fixation within the C1-C2 lateral mass articulation.

PURPOSE: To assess the atlantoaxial stability achieved with a novel integrated device when compared with the intact, destabilized, and stabilized state using the Harms technique.

STUDY DESIGN: A biomechanical study of implants in human cadaveric cervical spines.

METHODS: Six human cadaveric specimens were used. Biomechanical testing was performed with moment control in flexion-extension, lateral bending, and axial rotation. Range of motion (ROM) was measured in the intact state, after both destabilization by creation of a Type II odontoid peg fracture and sequential stabilization using the integrated device and the Harms technique.

RESULTS: Mean flexion-extension ROM of the intact specimens at C1-C2 was 14.1°±2.9°. Destabilization increased the ROM to 31.6°±4.6°. Instrumentation with the Harms technique reduced flexion-extension motion to 4.0°±1.4° (p<.01). The integrated device reduced flexion-extension motion to 3.6°±1.8° (p<.01). In lateral bending, the respective mean angular motions were 1.8°±1.1°, 14.1°±5.8°, 1.4°±0.7°, and 0.4°±0.3° for the intact destabilized Harms technique and integrated device. For axial rotation, the respective mean values were 67.3°±13.8°, 74.2°±16.1°, 1.4°±0.7° and 0.9°±0.7°. Both the Harms technique and integrated device significantly reduced motion compared with the destabilized spine in flexion-extension, lateral bending, and axial rotation (p<.05). Direct comparison of the Harms technique and the integrated device revealed no significant difference (p>.10).

CONCLUSIONS: The integrated device resulted in interference fixation at the C1-C2 lateral mass joints with comparable stability to the Harms technique. Perceived advantages with the integrated device include avoidance of fixation below the C2 lateral mass where the vertebral artery is susceptible to injury, and access to the C1 screw entry point through the blade of the integrated device avoiding extended dissection superior to the C2 nerve root and its surrounding venous plexus.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
22341395
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"