Clinical implication of right ventricular to left ventricular interlead sensed electrical delay in cardiac resynchronization therapy.

AIMS: To evaluate the clinical implication of right ventricular (RV) to left ventricular (LV) interlead sensed electrical delay (RV-LVs) and the relation to ventricular lead position in cardiac resynchronization therapy (CRT).

METHODS AND RESULTS: Eighty-five consecutive CRT patients (mean age 66 ± 11 years) received LV lead prospectively targeted to the latest mechanical activated segment (concordant), assessed by two-dimensional speckle tracking radial strain (ST-RS) echocardiography. The RV lead was randomized to RV apex (n= 43) or RV high posterior septum (n= 42). Right ventricular to left ventricular interlead sensed electrical delay was obtained during the CRT implant procedure. Intraventricular dyssynchrony was evaluated by ST-RS echocardiography. Interventricular mechanical delay (IVMD) was measured by using pulse-wave Doppler. Separated by the median RV-LVs (82 ms), a long RV-LVs demonstrated more LV end-systolic volume (LVESV) reduction than a short RV-LVs (-27 ± 20 vs. -16 ± 22%; P= 0.02), 6 months after CRT (6FU). Right ventricular to left ventricular interlead sensed electrical delay correlated to IVMD (r = 0.50; P< 0.001) and intraventricular dyssynchrony (r = 0.25; P= 0.02) at baseline. Concordant LV leads (n= 61) demonstrated superior reduction of LVESV (P= 0.005) 6 months after CRT; however, both RV lead positions had similar effects. Right ventricular to left ventricular interlead sensed electrical delay was irrespective to LV lead concordance and RV lead position (P= ns). Independent predictors to reverse remodelling (reduction of LVESV ≥ 15%) at 6FU were concordant LV lead (odds ratio, 3.210; P= 0.029) and IVMD (odds ratio, 1.028; P= 0.026).

CONCLUSION: Right ventricular to left ventricular interlead sensed electrical delay was not predictive to LV reverse remodelling affected by CRT at 6FU. Concordant LV leads demonstrated superior LV reverse remodelling at 6FU. Right ventricular to left ventricular interlead sensed electrical delay was irrespective of ventricular lead position and might be insufficient to target optimal LV lead position in CRT.

TRIAL REGISTRATION: https://clinicaltrials.gov. Unique identifier: NCT01035489.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app