Four-dimensional phase contrast MRI with accelerated dual velocity encoding.
PURPOSE: To validate a novel approach for accelerated four-dimensional phase contrast MR imaging (4D PC-MRI) with an extended range of velocity sensitivity.
MATERIALS AND METHODS: 4D PC-MRI data were acquired with a radially undersampled trajectory (PC-VIPR). A dual V(enc) (dV(enc) ) processing algorithm was implemented to investigate the potential for scan time savings while providing an improved velocity-to-noise ratio. Flow and velocity measurements were compared with a flow pump, conventional 2D PC MR, and single V(enc) 4D PC-MRI in the chest of 10 volunteers.
RESULTS: Phantom measurements showed excellent agreement between accelerated dV(enc) 4D PC-MRI and the pump flow rate (R(2) ≥ 0.97) with a three-fold increase in measured velocity-to-noise ratio (VNR) and a 5% increase in scan time. In volunteers, reasonable agreement was found when combining 100% of data acquired with V(enc) = 80 cm/s and 25% of the high V(enc) data, providing the VNR of a 80 cm/s acquisition with a wider velocity range of 160 cm/s at the expense of a 25% longer scan.
CONCLUSION: Accelerated dual V(enc) 4D PC-MRI was demonstrated in vitro and in vivo. This acquisition scheme is well suited for vascular territories with wide ranges of flow velocities such as congenital heart disease, the hepatic vasculature, and others.
MATERIALS AND METHODS: 4D PC-MRI data were acquired with a radially undersampled trajectory (PC-VIPR). A dual V(enc) (dV(enc) ) processing algorithm was implemented to investigate the potential for scan time savings while providing an improved velocity-to-noise ratio. Flow and velocity measurements were compared with a flow pump, conventional 2D PC MR, and single V(enc) 4D PC-MRI in the chest of 10 volunteers.
RESULTS: Phantom measurements showed excellent agreement between accelerated dV(enc) 4D PC-MRI and the pump flow rate (R(2) ≥ 0.97) with a three-fold increase in measured velocity-to-noise ratio (VNR) and a 5% increase in scan time. In volunteers, reasonable agreement was found when combining 100% of data acquired with V(enc) = 80 cm/s and 25% of the high V(enc) data, providing the VNR of a 80 cm/s acquisition with a wider velocity range of 160 cm/s at the expense of a 25% longer scan.
CONCLUSION: Accelerated dual V(enc) 4D PC-MRI was demonstrated in vitro and in vivo. This acquisition scheme is well suited for vascular territories with wide ranges of flow velocities such as congenital heart disease, the hepatic vasculature, and others.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app