The safety of energy-based devices in open thyroidectomy: a prospective, randomised study comparing the LigaSure™ (LF1212) and the Harmonic® FOCUS.
Langenbeck's Archives of Surgery 2012 June
PURPOSE: The options for haemostasis and dissection during modern thyroid surgery include the use of a bipolar vessel sealing system and ultrasonic technology. The aim of this study was to compare these energy-based devices (EBDs) with respect to their use in open thyroidectomy.
METHODS: The FOCUS Shear (F) and LigaSure LF1212 (L) were evaluated. A total of 182 consecutive patients scheduled for open thyroidectomy were prospectively randomised into two similarly sized groups. The parameters of operative time, morbidity, incision length, postoperative pain and hospital stay were analysed.
RESULTS: The main clinically primary endpoints of the two devices were identical in the study as the rate of nerve lesions, bleeding, drainage, operative time and postoperative calcium concentration with no significant differences with respect to the instrument utilised. The mean length of the incision was greater in the F group (p < 0.05). Patients in the F group were more likely to complain of pain while swallowing (p < 0.001). Early postoperative measurements of intact parathyroid hormone plasma levels revealed that although the levels were within the reference range, they were significantly lower in the F group (p < 0.001). Oral calcium supplementation was significantly higher and more prolonged in the F group.
CONCLUSIONS: The present study demonstrated no significant difference in the rates of postoperative morbidity associated with these two different EBDs used. Differences in clinically less significant were founded and focused on.
METHODS: The FOCUS Shear (F) and LigaSure LF1212 (L) were evaluated. A total of 182 consecutive patients scheduled for open thyroidectomy were prospectively randomised into two similarly sized groups. The parameters of operative time, morbidity, incision length, postoperative pain and hospital stay were analysed.
RESULTS: The main clinically primary endpoints of the two devices were identical in the study as the rate of nerve lesions, bleeding, drainage, operative time and postoperative calcium concentration with no significant differences with respect to the instrument utilised. The mean length of the incision was greater in the F group (p < 0.05). Patients in the F group were more likely to complain of pain while swallowing (p < 0.001). Early postoperative measurements of intact parathyroid hormone plasma levels revealed that although the levels were within the reference range, they were significantly lower in the F group (p < 0.001). Oral calcium supplementation was significantly higher and more prolonged in the F group.
CONCLUSIONS: The present study demonstrated no significant difference in the rates of postoperative morbidity associated with these two different EBDs used. Differences in clinically less significant were founded and focused on.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app