Add like
Add dislike
Add to saved papers

Fractured Bard Recovery, G2, and G2 express inferior vena cava filters: incidence, clinical consequences, and outcomes of removal attempts.

PURPOSE: To increase the understanding of risks of inferior vena cava (IVC) filter fracture and embolization and the safety of removing fractured filters via retrospective review of a prospectively collected database of fractured IVC filters.

MATERIALS AND METHODS: A total of 63 fractured IVC filters were discovered among 548 patients presenting for retrievable filter removal between April 2004 and November 2010 at a single institution. Device type, duration of implantation, component fracture, and embolization events were recorded. Success rates and techniques for removal of components were recorded.

RESULTS: A total of 63 fractured Recovery, G2, and G2 Express IVC filters were identified, for an overall fracture rate of 12%. Excluding foot process fractures, the fracture rate for only filter arms and/or legs was 6%. The incidence of fracture increased with longer filter dwell times. Success rates for removal of the nonfractured component (ie, main body) and fractured components (ie, arm or leg) were 98.4% and 53.4%, respectively. The distal embolization rate of fractured filter components was 13%. There were no immediate clinically significant complications associated with fracture component embolization or filter removal. A single patient was encountered with symptoms related to their fractured filter.

CONCLUSIONS: IVC filter fracture rates increase with longer dwell times; however, removal of fractured filters and fractured components (ie, arms and legs) can be achieved safely and effectively. Clinically significant complications of IVC filter fracture are rare, and there were no immediate clinical sequelae related to embolization of fracture components.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app