CLINICAL TRIAL
JOURNAL ARTICLE
MULTICENTER STUDY
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Prospective multicenter clinical trial (STABLE) on the endovascular treatment of complicated type B aortic dissection using a composite device design.

OBJECTIVE: This study evaluates the safety and effectiveness of a unique composite thoracic endovascular aneurysm repair (TEVAR) construct (proximal stent graft and distal bare metal stent) for the treatment of patients with complicated type B aortic dissection (cTBAD).

METHODS: In this prospective, single-arm, multicenter study, patients with cTBAD were treated with an endovascular system consisting of proximal TX2 thoracic stent grafts and distal bare metal dissection stents (Zenith Dissection Endovascular System; Cook Medical, Bloomington, Ind). Indications for enrollment were branch vessel malperfusion, impending rupture, aortic diameter ≥40 mm, rapid aortic expansion, and persistent pain or hypertension despite maximum medical therapy. One-year follow-up results, including clinical and radiographic (computerized tomography [CT] and X-ray) evaluation, were available for this report.

RESULTS: Ten centers enrolled 40 patients (70% men; mean age 58 years old) between December 2007 and August 2009. The onset of symptoms was acute (≤14 days) in 24 patients (60%), subacute (15-30 days) in six patients (15%), and chronic (31-90 days) in 10 patients (25%); the overall mean time from symptom onset to treatment was 20 days (range, 0-78 days). A majority of patients (77.5%; 31 of 40 patients) presented with impending aortic rupture (indicated by periaortic effusion/hematoma) or branch vessel malperfusion. Seven combinations of stent grafts and dissection stents were used, and all devices were successfully deployed and patent. The 30-day mortality rate was 5% (2 of 40); two deaths occurred after 30 days, leading to a 1-year survival rate of 90%. Two deaths, occurring at 11 and 81 days postprocedure, respectively, were secondary to aortic rupture. Morbidity occurring within 30 days included stroke (7.5%), transient ischemic attack (2.5%), paraplegia (2.5%), retrograde progression of dissection (5%), and renal failure (12.5%). Additional morbidity after 30 days included one case of retrograde progression of dissection and one case of renal failure. None of the patients with renal failure became dialysis-dependent. Four patients (10%) underwent secondary interventions within 1 year. Favorable aortic remodeling was observed during the course of follow-up, indicated by an increase in the true lumen size and a concomitant decrease in the false lumen size along the dissected aorta, with completely thrombosed thoracic false lumen observed in 31% of patients at 12 months as compared to 0% at baseline.

CONCLUSIONS: Initial data with a composite TEVAR construct have demonstrated favorable clinical and anatomic results. Continued enrollment and long-term data are needed to assess the overall effectiveness of this treatment strategy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app