Microwave ablation: results in ex vivo and in vivo porcine livers with 2450-MHz cooled-shaft antenna.

BACKGROUND: Imaging-guided thermal ablation using different energy sources continues to gain favor as a minimally invasive technique for the treatment of primary and metastatic hepatic malignant tumors. This study aimed to evaluate the performance of microwave ablation with 2450-MHz internally cooled-shaft antenna in ex vivo and in vivo porcine livers.

METHODS: All studies were animal care and ethics committee approved. Microwave ablation was performed using a noncooled or cooled-shaft antenna in 23 ex vivo (92 ablations) and eight in vivo (36 ablations) porcine livers. Diameters of the coagulation zone were observed on gross specimens. The coagulation diameters achieved in different microwave ablation parameter groups were compared. Curve estimation analysis was performed to characterize the relationship between applied power and treatment duration and coagulation diameter (including short-axis and long-axis diameter).

RESULTS: Coagulation zones were elliptical and an arrowed-shaped carbonization zone around the shaft was observed in all groups. But the antenna track was also coagulated in the noncooled-shaft antenna groups. In ex vivo livers, the short-axis diameter correlated with the power output in a quadratic curve fashion (R(2) = 0.95) by fixing ablation duration to 10 minutes, and correlated with the ablation duration in a logarithmic curve fashion (R(2) = 0.98) by fixing power output to 80 W. The short-axis reached a relative plateau within 25 minutes. In in vivo livers, short-axis diameter correlated with the coagulation duration in a sigmoidal curve fashion (60 W group R(2) = 0.76, 80 W group R(2) = 0.87), with a relative plateau achieved within 10 minutes for power settings of 60 W and 80 W.

CONCLUSIONS: The internally cooled microwave antenna may be advantageous to minimize collateral damage. The short-axis diameter enlargement has a plateau by fixing power output.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app