Cross-validation of MR elastography and ultrasound transient elastography in liver stiffness measurement: discrepancy in the results of cirrhotic liver.
Journal of Magnetic Resonance Imaging : JMRI 2012 March
PURPOSE: To evaluate individual differences in liver stiffness measurement using both MR elastography (MRE) and ultrasound transient elastography (UTE) in patients with chronic liver disease.
MATERIALS AND METHODS: This study included 80 patients with chronic liver disease who underwent both UTE and MRE. MRE and UTE were performed using a pneumatic driver (60 Hz) and an ultrasound probe with a vibrator (50 Hz), respectively. Liver stiffness data measured using the two techniques (μ(UTE) and μ(MRE) ) were compared with respect to shear modulus. The patients were subdivided into four quartiles on the basis of average of the μ(UTE) and μ(MRE) values for each patient.
RESULTS: The analysis of the 4 quartile groups revealed that μ(UTE) was significantly higher than μ(MRE) in the two most stiff liver groups: μ(UTE) versus μ(MRE) , 7.5 (1.2) versus 6.0 (0.72) kPa for the group with [μ(UTE) + μ(MRE) ]/2 of 5.6-8.0 kPa; 15.1(4.2) versus 6.7 (1.4) kPa for the group with >8.0 kPa. However, in the least stiff liver group (i.e., the group with [μ(UTE) + μ(MRE) ]/2 < 3.2 kPa), μ(UTE) was significantly lower than μ(MRE) .
CONCLUSION: The shear modulus measured by UTE and MRE are not equivalent, especially in patients with stiff livers.
MATERIALS AND METHODS: This study included 80 patients with chronic liver disease who underwent both UTE and MRE. MRE and UTE were performed using a pneumatic driver (60 Hz) and an ultrasound probe with a vibrator (50 Hz), respectively. Liver stiffness data measured using the two techniques (μ(UTE) and μ(MRE) ) were compared with respect to shear modulus. The patients were subdivided into four quartiles on the basis of average of the μ(UTE) and μ(MRE) values for each patient.
RESULTS: The analysis of the 4 quartile groups revealed that μ(UTE) was significantly higher than μ(MRE) in the two most stiff liver groups: μ(UTE) versus μ(MRE) , 7.5 (1.2) versus 6.0 (0.72) kPa for the group with [μ(UTE) + μ(MRE) ]/2 of 5.6-8.0 kPa; 15.1(4.2) versus 6.7 (1.4) kPa for the group with >8.0 kPa. However, in the least stiff liver group (i.e., the group with [μ(UTE) + μ(MRE) ]/2 < 3.2 kPa), μ(UTE) was significantly lower than μ(MRE) .
CONCLUSION: The shear modulus measured by UTE and MRE are not equivalent, especially in patients with stiff livers.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app