Comparison of all-inside meniscal repair devices with matched inside-out suture repair.

BACKGROUND: All-inside meniscal repairs are performed with increasing frequency because of the availability of newly developed devices. A comparison of their biomechanical characteristics may aid physicians in selecting a method of meniscal repair.

HYPOTHESIS: All-inside meniscal repairs will be superior to their inside-out controls in response to cyclic loading and load-to-failure testing.

STUDY DESIGN: Controlled laboratory study.

METHODS: Sixty-six bucket-handle tears in matched porcine menisci were repaired using the Ultra FasT-Fix, Meniscal Cinch, Ultrabraid No. 0, and FiberWire 2-0 sutures. Initial displacement, cyclic loading (100, 300, and 500 cycles), and load-to-failure testing were performed. The displacement, response to cyclic loading, and mode of failure were recorded. The stiffness was calculated.

RESULTS: The Meniscal Cinch demonstrated a significantly higher initial displacement than the other methods tested (P = .04). No significant difference was found among the methods in response to cyclic loading. The inside-out FiberWire repair demonstrated the highest load to failure (120.8 ± 23.5 N) and was significantly higher than both the Meniscal Cinch (64.8 ± 24.1 N, P < .001) and the Ultra FasT-Fix (88.3 ± 14.3 N, P = .002). It was not significantly higher than the inside-out Ultrabraid suture repair (98.8 ± 29.2 N). The inside-out FiberWire repair had the highest stiffness (28.7 ± 7.8 N/mm). It was significantly higher than the Meniscal Cinch (18.0 ± 8.8 N/mm, P = .01). The most common mode of failure in all methods was suture failure.

CONCLUSION: An inside-out suture repair affords surgeons the best overall biomechanical characteristics of the devices tested (initial displacement, response to cyclic loading, and load to failure). For an all-inside repair, the Ultra FasT-Fix reproduces the characteristics of its matched inside-out suture repair more closely than the Meniscal Cinch.

CLINICAL RELEVANCE: Inside-out sutures and all-inside devices have similar responses to cyclic loading.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app